Scattered radiation (general radiography)
200
180
160
140
120
100
80
60
40
20
0
0
• S1,S2 (wall stand): 100 kV; SID 110 cm; tube/detector centre height
140cm over floor
• S3: 100 kV; SID 100 cm; table height of 70 cm (standard working height)
• S4,S5: 100 kV; SID 100 cm; table height of 70 cm (standard working
height)
Figure 17: Measurement of scattered radiation in zones of occupancy (Sx)
For the diagram above a maximum throughput of 30 exposures/hour was
used. This complies with a throughput of 15 patients/hour with typically 2
exposures done per patient. The measurement results in the figure above refer
to one exposure.
Related Links
Significant zones of occupancy at the radiographic table
Significant zones of occupancy at the radiographic wall stand
0,1
0,2
Kerma [uGy]
DR 400 | Introduction to DR 400 | 41
S1,S2
S3
S4,S5
0,3
0,4
0,5
on page 38
on page 39
3231E EN 20210708 1530