234 | DR 400 | Product Information
Minimum room height
Tabletop attenuation equivalent mm
Aluminum
Weight
Fixed height radiographic table
Elevating radiographic table
X-ray tube stand column
X-ray tube stand arm
X-ray tube plus collimator (maxi-
mum weight)
Maximum load on the radiographic
table
Movement ranges
Transversal axis or y-axis move-
ment (back & front)
Vertical axis or z-axis movement
(up & down)
Longitudinal axis (x-axis) move-
ment (right & left)
Alpha axis rotation (Angle of the
X-ray tube)
Beta axis rotation (swivel of the
X-ray tube arm around the tube
stand axis)
Bucky horizontal movement in
the table
3231E EN 20210708 1530
245 cm
≤ 0.7
According to
DIN EN 60601-1-3 with 100kV and
HVL 3.6 mm Al
FDA 21 CFR § 1020.30 (n) with
100kV and HVL 3.6 mm Al
290 kg
350 kg
120 kg
25 kg
40 kg
320 kg
± 7 cm
33.5 cm to 180 cm from the floor
The movement range may vary depend-
ing on the type of X-ray tube.
131 cm
±110° with mechanical detents at 0°,
±45°, ±90°
±90° with mechanical detents at 0°,
±45°, ±90°
50 cm