Display
Error
Action
no Tube mA
28
B
tube > 70°C
29
B/C
to save data push 'M'
30
-
'NOT' signal
31
B
door open
32
C
exp. time > 6.3s
33
C
exp. time > 3.2s
34
B/C
exp. time < 2ms
35
B/C
AEC exposure break
36
C
dose too low after
37
50ms
C
pulse delay too long
38
B
exp. prepare timeout
39
C
device ready timeout
40
B
starter timeout
41
B
DR 400 | X-Ray Generator Mini Console | 199
Explanation
Tube current < 50% after 30 ms
Hood temperature > 70°C
-
'EMERGENCY' safety signal active
Door contact open
Exposure time > 6.3 sec
Exposure time > 3.2 sec (automatic expo-
sure control)
Exposure time < 2 msec (automatic expo-
sure control)
Exposure aborted by operator (automatic
exposure control)
Dose too small after 50 ms (automatic ex-
posure control)
Exposure pause between 2 pulses > 2 sec
(automatic exposure control)
Exposure preparation timeout
Device ready timeout
Normal speed starter timeout
3231E EN 20210708 1530