DR Detector is Exceeding the Maximum Working
Temperature
Details
Cause
Brief Solution
A message is displayed on NX indicating that the DR De-
tector is exceeding the maximum working temperature.
Due to ambient temperature conditions and the number
of acquired images, the DR Detector's internal tempera-
ture may become too high.
1. Power off the DR Detector.
2. Leave the DR Detector unpowered for at least one
hour.
3. Stop the NX workstation.
4. Power on the DR Detector.
5. Start the NX workstation.
DR 400 | Problem solving | 211
3231E EN 20210708 1530