168 | DR 400 | Bucky
Loading of the bucky in the Radiographic Table
To load the bucky with a cassette or detector:
1. Open the bucky drawer completely by pulling the front handle.
2. Push the cassette or detector towards the rear slider to open the clamping
mechanism wide enough to contain the cassette or detector.
3. Let the cassette or detector slip into the clamping.
4. Align the cassette or detector center indication to the center mark on the
clamp.
5. Close the bucky drawer using the front handle and pushing the button to
release the brake.
Make sure the bucky drawer is pushed up to the end to close completely.
Related Links
Orientation of DX-D 10C, DX-D 10G in the bucky
3231E EN 20210708 1530
CAUTION:
Make sure your fingers are not between slider and cassette.
The clamping mechanism may hurt your fingers, therefore
take special care.
CAUTION:
When positioning the cassette or detector out of center:
• The alignment to the X-ray tube must be controlled
manually.
• The AEC sensors might not be covered or not covered
completely, causing wrong exposure dose. Make sure
that AEC sensors are covered.
on page 183