Automatic collimation always too wide or too narrow

Details
Cause
Brief Solution
The collimated area is not adapted correctly to the size
of the cassette or DR Detected inserted in the bucky.
The sensors in the bucky that detect the size of the cas-
sette or DR Detector are dirty or have become weak.
Wipe the sensors in the bucky with a lint-free cloth. If
needed, moisten the cloth with a neutral detergent.
Figure 80: Location of the sensors in the bucky
If the problem persists, contact your local service or-
ganization to exchange the sensors.
DR 400 | Problem solving | 207