Grids

Anti-scatter grids are used to reduce scattered radiation and improve image
quality. Grids are available as an option.
For DR Detectors focused grids are used. Focused grids require centering of
the X-ray source to the detector and a specific distance range between X-ray
source and detector. The color of the handle of the grid indicates which
distance the grid is used for.
To change the grid in the radiographic table or radiographic wall stand:
1. Pull out the grid using the handle.
2. Store the grid in a safe place to avoid damage.
3. Insert the grid with labels facing up in the appropriate slit of the bucky.
Make sure the grid is pushed up to the end.
Related Links
Bucky Unit Technical Data
Topics:
Anti-scatter grids
Grid focal distance color indication
Grid detection
Storage box for DR Detector and grids
CAUTION:
Using a focused grid with the X-ray source not centered or on a
wrong distance may cause reduced image quality.
CAUTION:
Handle grids with care and store them in a safe place when not
in use. Dropping the grid can cause damage and create visible
image artifacts or reduce image quality.
CAUTION:
If grid is not inserted completely, artifacts on the image can be
visible, e.g. of the grid edges. Push the grid all the way up to the
end.
on page 240
DR 400 | Bucky | 185
3231E EN 20210708 1530