X-ray tube stand tracks table height

To keep constant SID while adjusting table height:
1. Set the required SID by adjusting the position of the X-ray tube stand.
The distance between the X-ray tube head and the table top must not be
less than 50 cm.
2. On the tube head display, press the position tracking button.
Figure 34: Table position tracking disabled and enabled
The button is highlighted.
3. Adjust the table height.
The X-ray tube stand is moving up or down accordingly.
The lock button controls the behaviour of the position tracking after the
exposure is made.
Table 6: Locking the position tracking
DR 400 | Software Console and Tube Head Display | 103
Note: The movement of the X-ray tube stand has a small delay
compared to the movement of the table. The movement of the X-
ray tube is automatically stopped if the distance between the X-
ray tube head and the table would become too small (SID lower
than 45 cm).
Position tracking is not active for the next exposure. It can
be activated again by pressing the position tracking but-
ton.
Position tracking is kept active for the next exposure.