134 | DR 400 | Radiographic Table and X-Ray Tube Stand

Stop positions

The system includes stop positions.
• On the longitudinal axis movement, to position the X-ray tube in regularly
used exposure distances to the radiographic wall stand, e.g. 150 cm and
180 cm.
• On the vertical axis movement, to position the X-ray tube stand in
regularly used exposure distances to the radiographic table, e.g. 115 cm.
The preferred positions of the stops are defined during installation.
The vertical stop on the tube stand is always active. Vertical stops on the tube
stand are not available on the fixed height radiographic table types TS-Fix-
L-001 and TS-Fix-R-001.
The two transversal stops are active when the X-ray tube is rotated towards
the wall stand (90°+/- 10°).
To enter a stop position, move the X-ray tube stand or the X-ray tube arm in
longitudinal or vertical direction. The movement is stopped when the stop
position is reached. Moving too fast may cause the X-ray tube stand to skip the
stop position.
To leave a stop position, release and press again the according movement
control button.
3231E EN 20210708 1530