210 | DR 400 | Problem solving

No table movement

Details
Cause
Brief Solution
3231E EN 20210708 1530
The table is not moving up or down when pressing the
foot pedals with double click. No error is shown.
One of the foot pedals was pressed longer than 90 sec-
onds.
1. Press the Power OFF button on the X-ray generator
control box to switch off the generator.
2. Switch off the electrical room switch.
3. Wait for 30 seconds.
4. Switch on the electrical room switch.
5. Press the Power ON button on the X-ray generator
control box to switch on the system.