Disinfecting safety directions

To disinfect the device:
• Do not use any corrosive, soluble or gaseous disinfectants.
• Consult the manufacturer's Material Safety Data Sheets (MSDS) and
recommendations on the product label for additional information prior to
use.
• Use of spray disinfection can cause malfunctions due to ingress of the
disinfectant into the equipment. Disinfect all parts of the unit, including
the accessories and connection cables by just wiping them. Switch off the
system and cover the cooled system carefully before performing a room
disinfection using nebuliser.
• Using unsuitable disinfectants can cause discoloration and damage of the
surface of the equipment.
3231E EN 20210708 1530
WARNING:
Using a disinfectant that can form an explosive or flammable
gas mixtures is hazard to life and health because of explosion
risk. Switch the equipment off before disinfecting. Allow the gas
mixture to evaporate before switching the x-ray system back on.