Emergency stop button

Figure 13: Emergency stop button
If a system malfunction causes an emergency situation involving the patient,
operating personnel or any system component, activate the emergency stop on
the radiographic table. All motor driven movements will be stopped.
Motor driven movements:
• Radiographic table
• Radiographic wall stand
• X-ray tube stand
To allow motorized movements again, turn the cap of the emergency switch in
clockwise direction (default position).
3231E EN 20210708 1530
WARNING:
The emergency stop button does not switch off the voltage in
the X-ray system.