68 | Full Leg Full Spine DR Retrofit System | Problem solving
Not all three DR detector positions are available
Table 8: Problem: Not all three DR detector positions are available
Details
Cause
Brief Solution
0326A EN 20211014 1141
After pushing the initialization button on the external
collimator, not all three leds light up.
If the top most led or the bottom most led doesn't light
up, this indicates that the external collimator cannot
reach that position and no exposure can be made in
that position.
The middle position of the external collimator is too
low or too high.
The examination can be continued if only two images
are needed. If an image is required for the missing posi-
tion, the workflow must be restarted with an adjusted
center position.