Performance Tests

Table 21 Equipment Required for Performance Tests

Description
Mainframe,
Agilent 8163A/B
Lightwave Multimeter
Device under Test
Laser Source
Agilent 81650A
(any other Agilent 8165x series
Laser Source)
Power Sensor
Agilent 81632B
(any other Agilent 8163x series
Power Sensor)
(Agilent 81618A or
Agilent 81619A Interface Mod-
ule plus
Agilent 81626B Optical Head or
any other Optical Head from the
Agilent 8162x series)
When using a high power laser source of type Agilent 81655A, Agilent 81656A, or Agilent 81657A,
N O T E
you must use an Interface Module and a high power Optical Head of type Agilent 81626B.
Tunable Laser Module N/A
Connector Adapter
2 x Agilent 81000AI
(depending on patchcord)
Singlemode Fiber
Agilent 81101AC
(other Agilent patchcord de-
pending on DUT's conenctor
option)
352
All manuals and user guides at all-guides.com

Equipment Required

Equipment required for the Performance Test is listed in the table below.
Any equipment that satisfies the critical specifications given in the table
may be substituted for recommended models.
Agilent 8164A/B
Lightwave Measurement Sys-
tem
Agilent 81650A
(any other Agilent 8165x series
Laser Source)
Agilent 81632B
(any other Agilent 8163x series
Power Sensor)
(Agilent 81618A or
Agilent 81619A Interface Mod-
ule plus
Agilent 81626B Optical Head or
any other Optical Head from the
Agilent 8162x series)
Agilent 81640A/B
(Agilent 81680A/B)
(Agilent 81682A/B)
2 x Agilent 81000AI
(depending on patchcord)
Agilent 81101AC
(other Agilent patchcord de-
pending on DUT's conenctor
option)
Part numbers in brackets indicate alternative models.
Model
Agilent 8166A/B
Lightwave Multichannel Sys-
tem
Agilent 81650A
(any other Agilent 8165x series
Laser Source)
Agilent 81632B
(any other Agilent 8163x series
Power Sensor)
(Agilent 81618A or
Agilent 81619A Interface Mod-
ule plus
Agilent 81626B Optical Head or
any other Optical Head from the
Agilent 8162x series)
N/A
2 x Agilent 81000AI
(depending on patchcord)
Agilent 81101AC
(other Agilent patchcord de-
pending on DUT's conenctor
option)
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition