Errors appearing on pop-up menus
error message
description
affected modules
action
error message
description
affected modules
action
error message
description
affected modules
action
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
All manuals and user guides at all-guides.com
Module communication error
There was an error in communication between module and mainframe. For
safety reasons, the module was disabled and the shutter closed.
All attenuators
If this problem occurs several times, please send the module to the local Agilent
service center for repair.
Offset table is empty
The user has tried to activate the Offset(Lambda) function, but the table of
wavelength dependent offsets is empty.
All attenuators
Please construct a table with at least one entry, before activating this mode.
Lambda offset active
The user has tried to program the table of wavelength dependent offsets, but
the Offset(Lambda) mode is activated.
All attenuators
Please deactivate the Offset(Lambda) mode first.
Error Messages
389