Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
All manuals and user guides at all-guides.com
E1x031: 0.25V ref out of limits (a, b, c)
The 0.25V reference voltage measured through the 16bit ADC is
out of the acceptable limits
a
b
c
x
Please return the unit to your local service center for repair
E10040: 10bit ADC test failed (a, b, c)
The 10bit ADC has returned a value other than 512
a
b
c
Please return the unit to your local service center for repair
E1x050: TEC NTC temp ou of limits (a, b, c, d)
The sensor temperature is out of the acceptable limits
a
b
c
d
x
Verify whether the module is operated within the specified ambi-
ent temperature. If the failure persists, please return the unit to
your local service center for repair.
E1x051: TEC current out of limits (a, b, c)
Verifies whether the TEC current is within a given range
a
b
c
x
Please return the unit to your local service center for repair
shows the ADC reading (0.25 = 3276)
shows the lower limit
shows the upper limit
refers to the module channel tested
shows the ADC reading
shows the lower limit
shows the upper limit
shows the sensor NTC temperature
shows the sensor NTC temperature ADC reading
shows the lower limit
shows the upper limit
refers to the module channel tested
shows the ADC reading of the TEC current
shows the lower limit
shows the upper limit
refers to the module channel tested