Fixed Laser Sources (Fabry Perot) - Agilent Technologies 8163A Manual

Lightwave multimeter
Hide thumbs Also See for 8163A:
Table of Contents

Advertisement

Fixed Laser Sources (Fabry Perot)

error message
description
test value
action
error message
description
test value
action
error message
description
test value
action
error message
description
test value
action
error message
description
test value
action
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
All manuals and user guides at all-guides.com
Fixed Laser Sources (Fabry
Perot)
E10010: Invalid Caldata
The Caldata loaded on the module is invalid. Either the caldata
version number is not compatible with the firmware or the calda-
ta is corrupt (checksum incorrect)
Please return the unit to your local service center for repair
E10012: Dual ported RAM write/read test failed
Writing bit patterns and reading them back to/from the dual port-
ed ram failed
Please return the unit to your local service center for repair
E10013: FPGA not correctly initialized
Loading the FPGA with data failed. This indicates a defective
FPGA or an FPGA access problem
Please return the unit to your local service center for repair
E10040: 10bit ADC test failed (a, b, c)
The 10bit ADC has returned a value other than 512
a
b
c
Please return the unit to your local service center for repair
E10050: Board temp out of limits (a)
The board temperature is outside the allowed range
a
Please return the unit to your local service center for repair
(no test value)
(no test value)
(no test value)
shows the ADC reading
shows the lower limit
shows the upper limit
measured board temperature
Error Messages
409

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

8163b8164a8164b8166a8166b

Table of Contents