Selftests

These tests are started after initialization of the instrument. This test measures
defined parameters, and tests them to limits.
Main Peltier Temp. Selftest Failed (error value)
Verifies whether the peltier temperature of the laser diode is within a given
range
13010
Peltier temperature is not in target range
Please return the unit to your local service center for repair
Main Thermo Temp. Selftest Failed (error value)
Verifies whether the temperature of the thermopositioner is within a given
range
13020
Thermopositioner is not in target temperature range
Please return the unit to your local service center for repair
Main Peltier Curr. Selftest Failed (error value)
Checks if peltier current for the laser diode peltier is below maximum
13050
Laser peltier current is above maximum value
Please return the unit to your local service center for repair
TCFS Temp. Selftest Failed (error value)
Verifies if the peltier temperature is within target with a given uncertainty
15010
TCFS peltier temperature is not in target range
Please return the unit to your local service center for repair
TCFS Resistor Selftest Failed (error value)
This test measures the R
15021
Photodiode R
15022
Photodiode R
15023
Photodiode R
Please return the unit to your local service center for repair
TCFS Attenuator Selftest Failed (error value)
Verifies whether an attenuator position is repeatable.
15030
Attenuator could not be positioned
Please return the unit to your local service center for repair
of the photodiodes at channel A, B and M
d
in channel A is below limit
d
in channel B is below limit
d
in channel C is below limit
d
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition