Figure 162 The Agilent 8163B Stability Analysis Screenn - First Screen - Agilent Technologies 8163A Manual

Lightwave multimeter
Hide thumbs Also See for 8163A:
Table of Contents

Advertisement

The Stability Application
NOTE
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
All manuals and user guides at all-guides.com

Figure 162 The Agilent 8163B Stability Analysis Screenn - First Screen

Setting Analysis Parameters
22To edit the power units used for the analysis, move to [W], press Enter, move to
<dBm>, <W>, or <dB>, and press Enter. See "What are the Power Units ?" on
page 85 for an explanation of power units.
23To edit the reference mode used for the analysis, move to <Ref Mode>, press
Enter, move to one of the following referencing modes:
• <Value>, you set a reference value using <Ref>, see step 24,
• <First Sample>, the first sample of the stability function is selected as the
reference value,
• <Channel 1>, the value that Channel 1 measures is updated as the reference
value for each set of power measurements (thus, each data value for
channel 1 will be 0 dB), and
• press Enter.
These settings have an effect when you choose <dB> as the <Pwr unit>.
See step 7 or step 22.
24If you have chosen <Value> as the <Ref Mode>, move to <Ref>, press Enter,
enter a value for the reference, and press Enter.