Tunable Laser Sources
error message
description
error value
action
error message
description
error value
action
error message
description
error value
action
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
All manuals and user guides at all-guides.com
Initialization Error (UpPiezoEnable, error value)
Over current
1
(not significant)
Please return the unit to your local service center for repair
Initialization Error - module temperature out of range
(please wait a little bit and try again)!
Module temperature is too high
(no error value)
Doublecheck whether the fan is working and not blocked. Wait some
time and reboot the instrument. If the failure persists, please return the
unit to your local service center for repair.
Initialization Error - module temperature too high
(please wait a little bit and try again)!
Module temperature is too high
(no error value)
Doublecheck whether the fan is working and not blocked. Wait some
time and reboot the instrument. If the failure persists, please return the
unit to your local service center for repair.
Error Messages