How to clean connector adapters

Some adapters have an anti-reflection coating on the back to reduce
back reflection. This coating is extremely sensitive to solvents and
mechanical abrasion. Extra care is needed when cleaning these
adapters.
Preferred Procedure
Use the following procedure on most occasions.
1 Clean the adapter by rubbing a new, dry cotton-swab over the surface using a
small circular movement.
2 Blow away any remaining lint with compressed air.
Procedure for Stubborn Dirt
Use this procedure particularly when there is greasy dirt on the adapter:
1 Moisten a new cotton-swab with isopropyl alcohol.
2 Clean the adapter by rubbing the cotton-swab over the surface using a small
circular movement.
3 Take a new, dry soft-tissue and remove the alcohol, dissolved sediment and
dust, by rubbing gently over the surface using a small circular movement.
4 Blow away any remaining lint with compressed air.

How to clean connector adapters

Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition