Screen Status Messages - Agilent Technologies 8163A Manual

Lightwave multimeter
Hide thumbs Also See for 8163A:
Table of Contents

Advertisement

Error Messages
Message
8163A/B
8164A/B
SL
SLock
Security Lock
RiL
RiLock
Remote InterLock
NoH
No Head
No Head
NoD
No Data
No Data
Set
Settling
Settling
NoR
No Ref
No Reference
B
Busy
Busy
382
All manuals and user guides at all-guides.com

Screen Status Messages

Meaning
The module is locked for security reasons. No measurement is possible
until you unlock the module
The remote interlock connector of your mainframe is open. No mea-
surement is possible until the connection at the remote interlock con-
nector closes
There is an optical head interface module inserted in the mainframe
that does not have an Optical Head connected to it. No measurement
is possible using this channel until you:
- reconnect an optical head or
- deselect this optical channel as a power measurement channel
No data has been measured or measurement is not finished. You can-
not access the analysis screen yet. Printout and file operations are not
yet possible.
A tunable laser source module is settling. The instrument is busy and
no measurement can be performed yet.
No reference sweep has been performed. Loss will be calculated using
nominal laser power as a reference.
Module is busy performing last operation.
Description
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition

Screen Status Messages