The Logging Application

Table 16 Logging Application On-Screen Messages

Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
All manuals and user guides at all-guides.com

On-Screen Messages

The messages listed in the table below may appear at the top of the screen
during the Logging application.
Agilent 8163A/B
Message
OvR
NoH
NoD
Agilent 8164A/B
Description
Message
Overrange
One of the Power Meter channels has re-
ported an overrange. The over-ranged
value is clipped, statistical data is invalid.
No Head
There is an Optical Head Interface mod-
ule inserted in the mainframe that does
not have an Optical Head connected to it.
No measurement is possible using this
channel until you either:
• reconnect an Optical Head or
• deselect this Optical Head channel as
a power measurement channel.
No Data
No data has been measured or measure-
ment is not yet finished. You cannot ac-
cess the Analysis Screen yet. Printout
and file operations are not yet possible.