Error Messages
error message
description
test value
action
408
All manuals and user guides at all-guides.com
E1x081: TEC current to temperature mismatch
If the TEC temperature difference does not correspond to the
measured Peltier current this test will fail
x
Verify whether the module is operated within the specified ambi-
ent temperature. If the failure persists, please return the unit to
your local service center for repair.