Error Messages
error message
description
test value
action
error message
description
test value
action
error message
description
test value
action
error message
description
test value
action
error message
description
test value
action
400
All manuals and user guides at all-guides.com
Return Loss Meters
E10010: Invalid Caldata
The Caldata loaded on the module is invalid. Either the caldata ver-
sion number is not compatible with the firmware or the caldata is
corrupt (checksum incorrect)
Please return the unit to your local service center for repair
E10012: Dual ported RAM write/read test failed
Writing bit patterns and reading them back to/from the dual ported
ram failed
Please return the unit to your local service center for repair
E10013: FPGA not correctly initialized
Loading the FPGA with data failed. This indicates a defective FPGA
or an FPGA access problem
Please return the unit to your local service center for repair
E10014: Firmware corrupt, checksum error (a,b)
Firmware in flash has an invalid checksum. Either the image which
was programmed was corrupt or the flash memory has a failure
a
b
Reload the module firmware. If the failure persists, please return the
unit to your local service center for repair
E1x031: 0.25 ref out of limits (a, b, c)
The 0.25V reference voltage measured through the 16bit ADC is out
of the acceptable limits
a
b
c
x
Please return the unit to your local service center for repair
(no parameter)
(no test value)
(no test value)
is the stored checksum
is the calculated checksum
shows the ADC reading (0.25V = 3276)
shows the lower limit
shows the upper limit
refers to the internal PM channel tested
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
Return Loss Meters