Mounting and installation

6
Mounting and installation
User qualification, mechanical instal‐
n
lation: trained qualified personnel, see
Ä Chapter 4.4 'Users' qualifications'
on page 21
User qualification, electrical installa‐
n
tion: Electrical technician, see
Ä Chapter 4.4 'Users' qualifications'
on page 21
24
NOTICE!
Mounting position and conditions
The controller fulfils the IP 67
degree of protection requirements
(wall mounting) or IP 54 (control
panel mounting) and NEMA 4X
(leak-tightness). These standards
are only fulfilled if all seals and
threaded connectors are correctly
fitted.
The installation (electrical) can
only take place after mounting
(mechanical)
Ensure that there is unimpeded
access for operation
Secure, low-vibration fixing
Avoid direct sunlight
Permissible ambient temperature
of the controller at the installation
location: -20 ... 60℃ at max. 95
% relative air humidity (non-con‐
densing)
The permissible ambient tem‐
perature of the connected sen‐
sors and other such components
must be considered
The controller is only suitable for
operation in closed rooms. If
operated outside, the controller
must be protected against the
environment by a suitable protec‐
tive enclosure