2-Point Fluoride Sensor Calibration (Cal) - ProMinent DULCOMETER Assembly And Operating Instructions Manual

Multi-parameter controller
Table of Contents

Advertisement

Calibrating the controller
2-point fluoride sensor cali‐
9.3.2
bration (CAL)
Correct sensor operation
Correct measuring and metering
is only possible if the sensor is
working perfectly
Observe the sensor operating
instructions
The carrying out of a 2-point cali‐
bration is strongly recommended
and is to be preferred to other
methods
For calibration the sensor must
be removed and refitted in the in-
line probe housing. To do this,
observe the operating instructions
of your in-line probe housing
Material required for calibration of fluoride
sensors:
Two test containers with calibrating
n
solution
78
Measuring and control behaviour
of the controller during calibration
During calibration the actuating out‐
puts are deactivated. Exception: a
basic load or a manual control vari‐
able has been set. This remains
active. The measured value output
[standard signal output mA] is frozen,
corresponding to its settings in the
mA output menu.
When calibration/testing has been
completed successfully, all of the
error checks relating to the measured
value are restarted. The controller
saves all the determined data for zero
point and slope upon a successful
calibration.
Used calibration solution
Dispose of the used calibration solu‐
tion. For more information: see cali‐
bration solution safety data sheet.
Two test containers with a calibration sol‐
ution are required for calibration. The fluo‐
ride content of the calibrating solutions
should be at least 0.5 ppm F
each other. The sensor should be rinsed
thoroughly with fluoride-free water when
changing the calibrating solution.
apart from
-