Disposal; Technical Data; Flow Modules - ProMinent DULCOMETER Assembly And Operating Instructions Manual

Multi-parameter controller
Table of Contents

Advertisement

Disposal/Technical Data
Page 30
8

Disposal

IMPORTANT
Please observe the applicable national regulations.
You may return the decontaminated used equipment to ProMinent Dosiertechnik
GmbH, Heidelberg, against prepaid postage.
9

Technical Data

IMPORTANT
• The maximum operating parameters for the whole in-line probe housing
(e.g. pressure, temperature, flow) are the lowest maximum operating
parameters of the in-line probe housing components and sensors
(please refer to the individual operating manuals)!
• Please also take the technical data of all other parts used, e.g. sensors, and
their operating manuals into account!
Storage and transport temperature
Humidity
9.1

Flow Modules

Weight
approx. 245 g (PG 13.5 module)
approx. 475 g (25 mm module)
Material
all modules
all fittings
seals
calibration cup PP
mounting panel white PVC
Temperature
max. 50 °C
Max. pressure
without flow sensor, 30 °C: 6 bar
without flow sensor, 50 °C: 1 bar
with flow sensor, 30 °C:
Flow rate
max. 80 l/h (40 l/h recommended)
Measuring accuracy
±5 l
of the flow module
Pressure loss in
flow module:
fitted modules
PG 13.5 module:
25 mm module:
-10 °C ... + 60 °C
with flow sensor: max. 90 % relative humidity,
non-condensing
transparent PVC
grey PVC
FPM
2 bar
12 mbar (12 cm WS)
2 mbar (2 cm WS)
20 mbar (20 cm WS)
ProMinent
®