Calibrating Fluoride (F ) Measured Value [Ma]; Selection Of The Calibration Process For Fluoride - ProMinent DULCOMETER Assembly And Operating Instructions Manual

Multi-parameter controller
Table of Contents

Advertisement

CAL ORP
Offset
Accept with <CAL>
Fig. 49: Acceptance of the [Offset]
5.
Press the
result of the calibration into the
memory of the controller
ð The controller displays the con‐
tinuous display again and oper‐
ates with the results of the cali‐
bration.
9.3
Calibrating fluoride (F
measured value [mA]
Selection of the calibration
9.3.1
process for fluoride
To calibrate the controller there are two
available calibration processes:
1 point
n
2 point
n
key to import the
-
)
Calibration process selection
1.
Continuous display ➨
CAL F -
1 ppm =
Slope
Single point calibration
A1034
Two point calibration
Fig. 50: Calibration menu [Fluoride]
ð The calibration menu is dis‐
played.
2.
Using the arrow keys select the
desired menu item. Press the
key
ð You can now start the selected
calibration process.
Calibrating the controller
185.0 mV
16:51:18
11/11/2011
-59.16 mV/dec
11:11:11
100 %
11/11/2011
A1037
77