Hydraulic Installation - ProMinent DULCOMETER Assembly And Operating Instructions Manual

Multi-parameter controller
Table of Contents

Advertisement

Figure 3
Installing the flow sensors
ProMinent
®
Flow sensor (see figure 3)
Remove the upper blanking plug of the flow module
Push the flow sensor (1) into the flow module
Tighten the reducing pipe nipple (3)
Tighten the clamping nipple (2)
1
2
3
80
60
40
25
l/h
Installing additional accessories:
Screw in the equipotential plug underneath the appropriate sensor
Instead of a lower blanking plug, screw in the sampling tap (two sizes: PG 13.5 or d = 25 mm).
5.2

Hydraulic Installation

IMPORTANT
• Observe the maximum operating parameters for the whole in-line probe
housing (e.g. pressure, temperature, flow)! Take into account the lowest
maximum operating parameters of the in-line probe housing components and
sensors (please refer to the individual operating manuals)! Please also note the
temperature dependence of the maximum pressure!
• Assemble the in-line probe housing in such a way that the modules cannot
drain off and fill with air, even when the water is stationary!
• There must be stop valves in the in-flow and out-flow of the module block!
• When assembling the in-line probe housing, take steps to prevent positive
suction pressure from building up inside it!
• When installing the in-line probe housing in a free flow system, lay the out-flow
cable in an ascending S-shape.
• Install a filter in the supply cable of the in-line probe housing if the water is
contaminated (take into account any depletion caused by the filter).
• If the in-line probe housing is installed in a fixed pipe, the power supply must
be switched off before the module block is fitted.
Connect the in-flow cable to the ball valve connector set
Connect the out-flow cable to the connector set on the out-flow side.
Assembly and Installation
Page 25

Advertisement

Table of Contents
loading

Table of Contents