Calibration process selection
1.
Continuous display ➨
CAL Cl
Last calibration 31.01. 2013 13:11:11
Zero point
4.00 mA
Slope
100 %
Slope calibration
Calibration of zero point
Fig. 56: Calibration menu [Chlorine]
ð The calibration menu is dis‐
played.
2.
Using the arrow keys select the
desired menu item. Press
ð You can now start the selected
calibration process.
9.4.2
Slope calibration
CAUTION!
Correct sensor operation / Run-in
period
Damage to the product or its sur‐
roundings
–
Correct measuring and metering
is only possible if the sensor is
working perfectly
A1039
–
Please read the operating manual
for the sensor
–
Please also read the operating
manuals for the fittings and other
components used
–
It is imperative that the run in
periods of the sensors are
adhered to
–
The run in periods should be
allowed for when planning com‐
missioning
–
It may take a whole working day
to run-in the sensor
Calibrating the controller