Disinfecting the control keys
G-30-1426-en
Maintenance / Further information
To be able to use the system in the OR, for example, it may be necessary
to disinfect the control keys. We recommend using MELISEPTOL disin-
fectant solution (B. Braun, Melsungen AG). Carl Zeiss keeps MELI-
SEPTOL in stock, and you can also obtain it locally in many countries
from representatives of B. Braun, Melsungen AG.
Warning!
– Wear disposable plastic gloves to prevent skin contact with the disin-
fectant.
– MELISEPTOL is inflammable (flame point at 31 °C). Please read the
product information from B. Braun, Melsungen AG.
To apply MELISEPTOL, proceed as follows:
•
Switch off the unit before disinfecting.
•
Apply the disinfectant across the entire surface of the control panel.
Do not let any disinfectant seep into the unit.
•
Leave the disinfectant on the unit for approx. 30 minutes.
•
Then carefully wipe the disinfectant off the surface using a sterile, lint-
free cloth.
•
Dispose of the gloves and cloth as normal waste.
You can obtain the following articles from Carl Zeiss:
1 l MELISEPTOL in vario bottle
MELISEPTOL HBV spray, 250 ml
MELISEPTOL HBV cloths
Disposable gloves:
Size 1 (large) size 8-9
Size 2 (medium) size 7-8
S88 Floor Stand with Lifting Column
157
Cat. No.
000000-0103-907
000000-0103-910
000000-0103-911
000000-0117-736
000000-0117-737
Issue 9.0