S705 - Adjust Hardware State From Control File - Zeiss IOLMaster 700 Service Manual

Hide thumbs Also See for IOLMaster 700:
Table of Contents

Advertisement

Adjustment and Calibration
S705 – Adjust hardware state from control file
Description:
Use this function to put the measurement
head and the PI into any desired state. The
desired hardware state can be specified in a
custom 'control-file'.
The use-case covers the following tasks:
(1) Select a pre-configured control-file.
(2) Parse and execute the control-file.
(3) Optionally, go one step back, select
another control-file and repeat the
previous actions, as often as required.
Instructions: (Step 1)
(1) Select the filename of the desired control-
file from the drop-down-box. NOTE: All
control files should be located in the
ControlFiles folder, if this is not the case it
will be necessary to select the folder.
(2) Click button 'Continue' to proceed or go
back to the use-case selection view.
Instructions: (Step 2)
(1) Wait, till the output-window indicates
that the control-file was executed.
(2) Check the 'Console-Output' for warnings
and errors.
(3) Click button 'Step back' to select another
control-file
OR
(4) Click button 'Finish' if you are done.
(Cleans up and returns to the use-case
selection view).
Service Manual
Order No. 2075-325
Fig. 21 – S705 Adjust hardware state from control file
(1)
Fig. 22 – S705 Adjust hardware state from control file
(2)
Fig. 23 – S705 Adjust hardware state from control file
(3)
Confidential and Proprietary
4 - 12
IOLMaster 700
SM-30-4060
Version 03, 2014-10-13