Zeiss S88 Instructions For Use Manual page 113

Floor stand with lifting column
Table of Contents

Advertisement

Configuration
mode 3
SETTINGS 3
EYE-
PIECE:
10x
SELECT
BRAKES:
ALL
AF-ROI
SIZE
SMALL
STORE MANUAL
MODE
G-30-1426-en
Operation
Configuration mode 3 for OPMI Vario (SETTINGS 3)
Configuration mode 3 for the OPMI Vario permits you:
– to enter the eyepiece magnification for the calculation of the total mag-
nification, menu item "EYE PIECE:"
– to select the magnetic brakes to be unlocked, menu item "SELECT
BRAKES:"
– to select the ROI size, menu item "AF-ROI SIZE"
This menu item is only displayed if the autofocus option has been in-
stalled.
Path: Every time you switch on the system, the basic mode is automati-
cally displayed.
To access configuration mode 3 from the basic mode, go via con-
figuration modes 1 and 2 by simultaneously pressing the "MODE"
and "STORE" buttons three times successively.
Entering the eyepiece magnification for the calculation of the total magni-
fication
In the basic mode, you can display the total magnification (MAG) of the
optical system as a factor in the middle display and key field. The total
magnification is calculated by the software. For this purpose, the magnifi-
cation of the eyepieces currently used must be known to the software.
Use menu item "EYE PIECE:" to select one of the eyepiece magnifica-
tions available.
Roll-over procedure:
S88 Floor Stand with Lifting Column
→ 16x → 10x →
12,5x
113
Issue 9.0
Printed on 28. 01. 2009