Failure of lamp control

Failure of focusing system

Failure of magnetic brakes

G-30-1426-en
Operation
Press Manual button (1) if brightness control is no longer possible.
Note:
When the manual function has been activated, all electrical control sys-
tems are inoperative. The lamp brightness is automatically adjusted to a
fixed setting.
Press Manual key (1) if, for example, the focusing system always
moves into its upper or lower end position.
Focus by moving the suspension arm of the ceiling mount (floor stand)
up or down.
If the magnetic brakes fail (magnetic brakes are locked), you can manu-
ally position the articulated arm including the microscope by overcoming
the locking effect of the magnetic brakes.