Connector panel

G-30-1426-en
Description
1 Remote control socket
for triggering an AUX signal, e.g. to switch on/off an external device
operating at max. 24V/0.5A.
2 Connector for switching component
Connection possibility for: a foot control panel, a hand control panel or
an operating chair with an appropriate foot switch.
3 Potential equalization bolt
4 Indicator window for rated voltage
The voltage shown here must correspond to the rated line voltage pro-
vided on the site of installation. You can adjust the sliding switch using
a suitable tool.
Warning!
Please observe the maximum current consumption of two power outlets
(4) and (5). Only connect medical devices approved by us to outlets (4)
and (5). When using other instruments, make sure that safety is guaran-
teed regarding admissible ground leakage currents. The admissible limit
value of the ground leakage current present in the suspension system's
power cord must not exceed 500 µA in compliance with EN60601-1/IEC
60601-1. CSA/US certification in compliance with UL 60601-1 only allows
a maximum ground leakage current of 300 µA.
5 Power outlet socket
for medical devices with a maximum current consumption of 2 A.
Note:
The current of this power outlet is switched on/off using power
switch S2 (7).
6 Power outlet socket
for medical devices with a maximum current consumption of 5 A.
7 Power inlet socket
8 Power switch S2
When the stand is on, the green indicator light in the switch is lit.
S88 Floor Stand with Lifting Column
Issue 9.0
Printed on 28. 01. 2009