Technical data
Electrical data of the stand
Power requirements Only connect the suspension system to wall out-
Rated voltage
Current
consump-
tion
Rated frequency
Fuses
Electrical
outlets
Electrical
standard
Approval
EMC requirements
The system has been designed for continuous operation.
S88 Floor Stand with Lifting Column
lets which are provided with a properly connected
protective ground conductor.
115 VΑC (100...125 VAC± 10%)
230 VAC (220...240 VAC± 10%)
115 VAC max.10 A
230 VAC max. 8 A
50...60 Hz
Automatic circuit breaker
– Power outlet 115/230 VAC, max. 5 A
– Power outlet 115/230 VAC, max. 2 A,
– (Via power switch),
– X-Y coupling
– Surgical microscope
– Remote control socket for an external signal
of a maximum of 24 V / 0.5 A.
Complying with IEC 60601-1 / EN 60601-1;
UL 60601-1;
CAN/CSA-C22.2 No. 601.1-M90
Protection class I,
degree of protection IPX0,
type B equipment,
product classification I in acc. with 93/42/EEC
Annex IX
The system meets the EMC requirements of
IEC 60601-1-2.
The system meets the RFI requirements of Class
A (hospital grade).
Issue 9.0
Printed on 28. 01. 2009