Configuration
mode 4
SETTINGS 4
AF-ROI
X-POS.
0
AF-ROI
Y-POS.
0
AF
V-FOC.
85
MODE
STORE MANUAL
-5
0
+5
-5
0
+5
Configuration
mode 4
SETTINGS 4
AF-ROI
X-POS.
0
AF-ROI
Y-POS.
0
AF
V-FOC.
85
MODE
STORE MANUAL
G-30-1426-en
Operation
Configuration mode 4 for OPMI Vario (SETTINGS 4)
Configuration mode 4 for the OPMI Vario permits you:
– to define the horizontal and vertical ROI position,
menu items "AF-ROI X-POS." and "AF-ROI Y-POS."
– to enter the video focal length, menu item "AF V-FOC"
Path: Every time you switch on the system, the basic mode is automati-
cally displayed.
To access configuration mode 4 from the basic mode, go via con-
figuration modes 1, 2 and 3 by simultaneously pressing the
"MODE" and "STORE" buttons four times successively.
Defining the position of the ROI
The two menu items "AF - ROI X- POS." and "AF - ROI Y- POS." permit
you to define the position of the ROI within the field of view of the micro-
scope.
When you start your system, the ROI is always positioned at the center of
the field of view. The ROI is visualized as a square in the graphic display.
Starting from the center position in the field of view, the ROI can be moved
in all directions by a maximum of 5 steps.
Use the arrow keys to shift the ROI:
– to the left or right using menu item "AF-ROI X-Pos."
– up or down using menu item "AF-ROI Y-Pos."
Entering the video focal length, menu item "AF-V-FOC"
Note:
This setting is performed during the installation of the SpeedFokus (auto-
focus) option. It is a camera-specific setting which only needs to be
changed when a different video camera is installed.
The video focal length of the video camera currently used must be known
to the SpeedFokus (autofocus) option. This menu item is used to specify
this camera-specific value to the SpeedFokus (autofocus) option.
This setting must always conform to the technical specifications of the
video camera used. The SpeedFokus (autofocus) option does not func-
tion properly if parameters have not been correctly set.
You can select the relevant focal length from the values offered.