Zeiss OPMI VISU 160 Instructions For Use Manual

Zeiss OPMI VISU 160 Instructions For Use Manual

On suspension systems
Table of Contents

Advertisement

OPMI
VISU 160
®
on S8, S81 & S88
Suspension Systems
Instructions for use
G-30-1529-en
Issue 5.0
Printed on 09. 10. 2006

Advertisement

Table of Contents
loading

Summary of Contents for Zeiss OPMI VISU 160

  • Page 1 OPMI VISU 160 ® on S8, S81 & S88 Suspension Systems Instructions for use G-30-1529-en Issue 5.0 Printed on 09. 10. 2006...
  • Page 2: Key To Symbols

    ® OPMI is a registered trademark of Carl Zeiss Surgical GmbH ™ Invertertube is a trademark of Carl Zeiss Surgical GmbH. ® Superlux is a registered trademark of Carl Zeiss Surgical GmbH G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0...
  • Page 3: Table Of Contents

    Contents – Key to symbols Functions at a glance – VISU 160 surgical microscope – Illumination systems – S88 floor stand – S8 ceiling mount – S81 ceiling mount Safety – Notes on installation and use – When using a wide-angle observation system (e.g. BIOM 3) –...
  • Page 4 – Intended use – Description of the modules – Design – Stand base with column – Connection panel – Instrument tray (option) S8 ceiling mount – Intended use – Description of the modules – Design – Power switch with connector (option) S81 ceiling mount –...
  • Page 5 Connections – Connecting the surgical microscope – Connecting the S light guide – Strain relief device on S88 floor stand – Connecting the S88 floor stand – Relocating the system Adjusting the supension system – Adjusting the balance setting of the suspension arm –...
  • Page 6 – Changing the halogen lamp – Changing the Superlux Eye xenon lamp module – Magnifications / Fields of view – Care of the unit – Sterilization – Disinfecting the control keys – Ordering data – Spare parts – Accessories – Disposal Technical data –...
  • Page 7: Functions At A Glance

    Functions at a glance Functions at a glance VISU 160 surgical microscope Illumination systems S88 floor stand S8 ceiling mount S81 ceiling mount G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 8: Visu 160 Surgical Microscope

    Functions at a glance VISU 160 surgical microscope Resetting the X-Y coupling and focus to their initial page 52 positions Tilting the surgical microscope page 148 Adjusting the eyecups page 70 Setting your prescription page 70 Handgrip page 62 Display of the magnification factor of the zoom system page 62 DeepView button page 64...
  • Page 9: Illumination Systems

    Functions at a glance Illumination systems Halogen illumination system Closed flap: main lamp is on - page 36 Open flap: backup lamp is on Selecting a filter page 36 Opening the lamp module page 36 Manual activation of the backup lamp page 36 Superlux Eye illumination system Selecting a filter...
  • Page 10 Functions at a glance Superlux Eye illumination system with additional integrated halogen illumination (option) Selecting the filter for Superlux Eye illumination page 82 Manual activation of the Superlux Eye backup lamp page 82 Opening the Superlux Eye lamp module page 74 Additional integrated halogen illumination: page 84 - Closed flap: main lamp is on...
  • Page 11 Functions at a glance G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 12: S88 Floor Stand

    Functions at a glance S88 floor stand Control panel page 158 Illumination system page 72 Releasing the magnetic brakes of the suspension page 88 system Limiting the suspension arm's downward movement page 142 Removing/mounting the coupling for the page 88 surgical microscope Balancing the suspension arm page 140...
  • Page 13 Functions at a glance 8 9 10, 11, 12 G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 14: S8 Ceiling Mount

    Functions at a glance S8 ceiling mount Locking the suspension arm in its horizontal position page 88 Releasing the magnetic brakes of the suspension page 88 system Balancing the suspension arm page 140 Illumination system page 72 Control panel (rotatable through 180° or 70°) page 158 Connecting the foot control panel or hand switch, page 106...
  • Page 15 Functions at a glance G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 16: S81 Ceiling Mount

    Functions at a glance S81 ceiling mount Locking the suspension arm in its horizontal position page 88 Releasing the magnetic brakes of the suspension page 88 system Balancing the suspension arm page 140 Illumination system page 72 Control panel (rotatable through 180° or 70°) page 158 Option: hand control panel socket for ceiling track mount page 112 Connecting the remote control connector...
  • Page 17 Functions at a glance G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 18 Functions at a glance G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 19: Safety

    Safety Safety Notes on installation and use When using a wide-angle observation system (e.g. BIOM 3) Phototoxic retinal injury in eye surgery Safety devices of the suspension systems Warning labels and notes G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09.
  • Page 20 Safety The device described in this manual has been designed and tested in ac- cordance with Carl Zeiss safety standards as well as German and inter- national standards. This guarantees a high degree of instrument safety. The system described in this user manual has been designed in compli- ance with the requirements of: –...
  • Page 21: Notes On Installation And Use

    • Over longer distances (e.g. removal, return for repair, etc), the instru- ment may only be transported in the original packaging or in special return packaging. Please contact your dealer or the Carl Zeiss service team. • Use this instrument only for the applications described.
  • Page 22 Safety • Only use the instrument with the accessories supplied. Should you wish to use other accessory equipment, make sure that Carl Zeiss or the equipment manufacturer has certified that its use will not impair the safety of instrument. •...
  • Page 23 – Ensuring that the navigation system is only connected by person- nel who have undergone appropriate training and instruction. – Contacting the local Carl Zeiss representative for any inquiries that may arise. – Implementation of a procedure that guarantees the calibration of the surgical microscope which is absolutely vital for the use of the Carl Zeiss components "Surgical microscope on suspension sys-...
  • Page 24 • The antenna manufacturer must confirm that his antenna has been tested and certified for operation with the respective Carl Zeiss surgi- cal microscope on a suspension system in accordance with the re- quirements specified in the Carl Zeiss interface description "Navigation Interface for Carl Zeiss Surgical Microscopes".
  • Page 25 Safety Requirements for operation • For ceiling mounts only: Our service staff or a qualified person ap- pointed by us will install the system on ceiling anchors which have been properly mounted by the construction engineers responsible. These ceiling anchors must comply with the specifications contained in our planning manual.
  • Page 26 Xenon lamps feature high luminance and a spectrum resembling that of natural daylight. Therefore, only special xenon lamps approved by Carl Zeiss must be used in ophthalmology. • Any kind of radiation has a detrimental effect on biological tissue.This also applies to the light illuminating the surgical field.
  • Page 27: When Using A Wide-Angle Observation System (E.g. Biom 3)

    Safety When using a wide-angle observation system (e.g. BIOM 3) When using a wide-angle observation system (e.g. BIOM 3 from Oculus) which is usually installed between the surgical microscope and the pa- tient, make sure that the patient is neither put at risk nor injured by the mo- torized focusing system or the movement of the suspension system arm.
  • Page 28 Safety In the following, comments on these aspects are given and a description of how Carl Zeiss, as a manufacturer, makes allowance for them in its systems. Illumination characteristics (spectral composition) Studies on exposure of the eye to light of varying spectral composition date back to the early 1950s.
  • Page 29 Also, the surrounding light sources should not cause addi- tional strain to the patient's eye. Carl Zeiss has provided an answer to this problem in the form of a swing- in retinal protection device for insertion into the beam path of the surgical microscope.
  • Page 30 Safety In conclusion Carl Zeiss recommends: – Use of the GG 475 retina protection filter. – Reduction of the illumination of the surgical area to the extent required for the patient's safety and for clear microscopic visualization. – Tilting of the microscope body as required.
  • Page 31 Safety G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 32: Safety Devices Of The Suspension Systems

    Safety Safety devices of the suspension systems 1 Release bar Allows non-sterile persons to release the magnetic brakes of the sus- pension system. 2 Adjustment screw for limiting the downward travel Use this screw to set the minimum vertical distance (working distance) from the surgical field.
  • Page 33 Safety G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 34 Safety Superlux Eye illumination system Warning! The xenon lamp has a limited service life of 500 h. If used beyond its maximum service life, the xenon lamp may explode. Change the xenon lamp in good time. 1 Switching to the backup lamp The lamp module contains two xenon lamps.
  • Page 35 Safety G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 36 Safety Halogen illumination system 1 Flap The flap is the mechanical indicator for the operating status of the hal- ogen lamps. – When the flap is closed, the main lamp is operative. – When the flap is open, the main lamp has failed. The backup lamp is on.
  • Page 37 Safety G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 38 Safety Superlux Eye illumination system with additional integrated halogen illumination (option) Warning! The xenon lamp has a limited service life of 500 h. If used beyond its maximum service life, the xenon lamp may explode. Change the xenon lamp in good time. 1 Flap The flap is the mechanical indicator for the operating status of the hal- ogen lamps.
  • Page 39 Safety G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 40 Safety Manual function 1 Manual key The Manual key permits you to switch to manual operation. The mo- torized control functions of the surgical microscope are deactivated. The lamp brightness is automatically adjusted to a fixed setting, the value being shown in the first display section. When the manual mode is activated, the yellow LED is lit and the word "MANUAL"...
  • Page 41 Safety G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 42: Warning Labels And Notes

    If any label is missing on your instrument or has become illegible, please contact us or one of our authorized representatives. We will supply the missing labels. OPMI VISU 160 G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0...
  • Page 43 Safety Illumination systems Superlux Eye GG 475 Superlux Eye GG 475 G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 44 Safety S88 floor stand G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 45 Safety S88 floor stand with instrument tray option G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 46 Safety S8 ceiling mount G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 47 Safety S81 ceiling mount G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 48 Safety G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 49: Description

    Description Description VISU 160 surgical microscope Intended use Description of the modules Illumination system Controls, displays, connections Binocular tubes and eyepieces Illumination systems Halogen illumination system Superlux Eye illumination system Superlux Eye illumination system with additional integrated halogen illumination (option) Identical modules of the suspension systems Suspension arm Display field with control keys...
  • Page 50 Description S81 ceiling mount Intended use Description of the modules Design Power switch, connector and socket (option) VISU 160 surgical microscope on S88 floor stand Intended use Design VISU 160 surgical microscope on S8 ceiling mount Intended use Design VISU 160 surgical microscope on S81 ceiling mount Intended use Design Foot control panel (option)
  • Page 51 Description G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 52: Visu 160 Surgical Microscope

    Description VISU 160 surgical microscope Intended use The VISU 160 surgical microscope has been designed for the magnified visualization of the field of view during surgical procedures in ophthal- mology. Description of the modules The VISU 160 surgical microscope is comprised of the following modules: 1 X-Y coupling The X-Y coupling allows motorized fine positioning of the surgical mi- croscope in a horizontal plane.
  • Page 53 Description G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 54 Description 2 Support arm for the surgical microscope The support arm incorporates a tilt device. This allows the viewing di- rection of the surgical microscope to be adapted to the requirements of the surgical field. Using the knob for fine tilt, you can position the surgical microscope in a range from +180°...
  • Page 55 Description G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 56 Description G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 57: Illumination System

    Description Illumination system The illumination system has been designed for use in ophthalmology. A light guide directs the light from the light source in the suspension system to the surgical microscope. A retinal protection device is provided to protect the patient's eye from photoretinitis.
  • Page 58 Description 6° illumination The 6° illumination can be faded out continuously. The result is a signifi- cantly reduced illumination reflex on the cornea. Despite this, the image provides high contrast and a high information content. A slight opening (approx. 1/4) of the 6° illumination is ideal for cataract surgery.
  • Page 59 Description + 6° + 2° G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 60: Controls, Displays, Connections

    Description Controls, displays, connections 1 Securing screw 2 X-Y coupling 3 Actuator button – recenters the X-Y coupling. – resets the focus to its initial position in the focusing range Note: Press this button to start the recentering movement. To stop this movement, press the button again.
  • Page 61 Description G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 62 Description 8 Dust cover 9 Handgrips with turning function for the release of the magnetic brakes of the suspension system (function only available on suspension sys- tems with magnetic brakes) For as long as you turn the left and/or right handgrip to the left or right and keep them in their turned position (turning function), the magnetic brakes of the suspension system are released.
  • Page 63 Description G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 64 Description 12 DeepView button Permits the selection of different modes. When you switch on the system, the mode for transmission optimization is set by default. This mode is recommended for viewing the posterior segment of the eye. Press the button to activate the mode for depth of field optimization. The green LED in the button is lit.
  • Page 65 Description G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 66: Binocular Tubes And Eyepieces

    Description Binocular tubes and eyepieces You can mount a 180° tiltable tube, an Invertertube or a 45° inclined tube on the VISU 160 surgical microscope as required (see the following pages). 180° tiltable tube 1 PD adjustment knob The correct position has been reached when the two eyepiece images merge into one.
  • Page 67 Description G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 68 Description Invertertube (option) The tiltable tube has an inverter function and has been designed for oph- thalmic use. Many wide-angle observation systems for the posterior seg- ment of the eye provide an inverted intermediate image which is viewed through the microscope. The inverter is used to erect an inverted image. When activating the wide-angle observation system, you must also acti- vate the inverter of the tiltable tube.
  • Page 69 Description G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 70 Description Widefield eyepieces with magnetic coupling Note: When you remove these eyepieces from the tube, please note that they are fitted with a magnetic coupling. When mounted, the eyepieces display a very weak magnetic field, so that the usual rules for the handling of mag- nets must only be observed with eyepieces which have not been mounted on the microscope: •...
  • Page 71 Description G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 72: Illumination Systems

    Warning! When using the Superlux Eye illumination system, only operate the system with special xenon lamps approved by Carl Zeiss for ophthalmic surgery. If any other than Carl Zeiss-approved xenon lamps are used, there is the risk of severe injury to the patient's eye.
  • Page 73 Description G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 74: Halogen Illumination System

    Description Halogen illumination system The ceiling mount is equipped with an illumination system for fiber illumi- nation. The lamp housing contains a backup lamp which is automatically swung into the illumination beam path when the first lamp fails. If required, the illumination system can be equipped with a second lamp housing so that two separate illumination systems are available for fiber illumination.
  • Page 75 Description G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 76 Description 7 Brightness control Brightness can be adjusted using the two keys (7) on the control panel. Note: If the suspension system has two lamp housings, you can also adjust the brightness of lamp 1 by pressing the appropriate button on the foot control panel.
  • Page 77 Description G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 78: Superlux Eye Illumination System

    Description Superlux Eye illumination system Warning! The xenon lamp has a limited service life of 500 h. If used beyond its maximum service life, the xenon lamp may explode. Change the xenon lamp in good time. The suspension system is equipped with a xenon illumination system for fiber illumination.
  • Page 79 Description When you press this button, the lamp module is slightly ejected. For changing the lamp, pull out the lamp module as far as it will go. Turn knob (2) through 180° until it snaps in. This moves the backup lamp into the illumination beam path.
  • Page 80 Description 6 Brightness control You can adjust the brightness using the two control keys on the control panel. Note: The brightness of the xenon lamp can also be adjusted by pressing the appropriate buttons on the foot control panel. 7 Yellow indicator lamp Lights when the lamp has failed, or if the lamp module is defective.
  • Page 81 Description G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 82: Superlux Eye Illumination System With Additional Integrated Halogen Illumination (Option)

    Description Superlux Eye illumination system with additional integrated halogen illumination (option) Warning! The xenon lamp has a limited service life of 500 h. If used beyond its maximum service life, the xenon lamp may explode. Change the xenon lamp in good time. The suspension system is equipped with a xenon illumination system for fiber illumination.
  • Page 83 Description 3 Opening the Superlux Eye lamp module When you press this button, the lamp module is slightly ejected. For changing the lamp, pull out the lamp module as far as it will go. Turn knob (2) through 180° until it snaps in. This moves the backup lamp into the illumination beam path.
  • Page 84 Description 4 Flap The flap is the mechanical indicator for the operating status of the hal- ogen lamps. – When the flap is closed, the main lamp is operative (green light (13) is on). – When the flap is open, the main lamp has failed. The backup lamp is operative (yellow light (12) is on).
  • Page 85 Description G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 86 Description 12 Yellow indicator lamp – Lights when the main lamp has failed. The backup lamp is on. – Blinks when the backup lamp has failed. 13 Green indicator lamp Indicates which illumination system is on. 14 Selector: Illumination is off. Illumination is on.
  • Page 87 Description G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 88: Identical Modules Of The Suspension Systems

    Description Identical modules of the suspension systems Suspension arm 1 Lock of the cable duct – For opening, turn a quarter turn clockwise or counterclockwise. – For closing, press down and turn a quarter turn clockwise or coun- terclockwise. 2 Adjustment screw for limiting downward movement Use this screw to set the minimum vertical working distance from the surgical field.
  • Page 89 Description G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 90: Display Field With Control Keys

    Description Display field with control keys The display and control panel is integrated in the control unit. Basic mode The surgical microscope on the suspension system can be controlled ei- Halogen ther manually or electronically. The control software required for elec- tronic control is installed in the electronics box of the suspension system.
  • Page 91 Description Keys Three keys and an LED are provided below the displays. Use the "MODE", "STORE" and "MANUAL" keys to select the different control functions (modes). "MODE" key and "STORE" key The "MODE" and "STORE" keys permit you to access the different modes of the user interface.
  • Page 92: S88 Floor Stand

    S88 floor stand Intended use The floor stand is a suspension system for Zeiss surgical microscopes suitable for almost all surgical disciplines. It is used to power and control the motorized functions of the surgical microscope. The hallmarks of the floor stand are its superb mobility and easy operation.
  • Page 93: Description Of The Modules

    Description Description of the modules The floor stand comprises an articulated arm, a stand column and a stand base. The articulated arm comprises a carrier arm and a suspension arm. The carrier arm contains the control unit with all electrical supply systems required for the control of a motorized surgical microscope.
  • Page 94: Design

    Description Design 1 Control panel 2 Carrier arm 3 Illumination system, see page 72 4 Suspension arm 5 Stand base G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 95 Description G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 96: Stand Base With Column

    Description Stand base with column 1 Maneuvering handle for moving the stand. 2 Support for hanging up the foot control panel during transport. 3 Cable support (2x) for winding up the power cord and the cable of the foot control panel. 4 Locking pedal Press once to lock the stand in position.
  • Page 97 Description G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 98: Connection Panel

    Description Connection panel 1 Remote control socket for triggering an AUX signal, e.g. to switch on/off an external device operating at max. 24V/0.5A. 2 Connector for switching component Possibility of connecting a foot control panel, hand control panel or op- erating chair with an appropriate footswitch.
  • Page 99 Description – connecting cable for foot control panel, hand control panel or op- erating chair with an appropriate footswitch. G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 100: Instrument Tray (Option)

    The instrument tray (1) can carry a maximum of 13 kg. It is designed for mounting the Zeiss instruments VISULAS 532s or MediLive Trio: – The four receptacles (2) on the instrument tray are intended for VISU- LAS 532s.
  • Page 101 Description G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 102: S8 Ceiling Mount

    S8 ceiling mount Intended use The S8 ceiling mount is a suspension system for Zeiss surgical micro- scopes. It is used to power and control the motorized functions of the sur- gical microscope. The hallmarks of the S8 ceiling mount are its superb mobility and easy operation.
  • Page 103: Description Of The Modules

    Description Description of the modules The S8 ceiling mount comprises an articulated arm, a suspension arm with the lamp housing and a control unit. The articulated arm consists of a lift arm and a carrier arm. The lifting func- tion permits the ceiling mount to be moved to a standby position. A grip is provided for height adjustment of the ceiling mount.
  • Page 104: Design

    Description Design 1 Lift arm 2 Carrier arm 3 Suspension arm 4 Grip for moving the ceiling mount into the standby or working position. 5 Illumination system, see page 72 6 Control panel G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09.
  • Page 105 Description G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 106: Power Switch With Connector (Option)

    Description Power switch with connector (option) The power switch and the connector can be either installed in the OR, or they can be integrated in the ceiling mount, at the back of the carrier arm (see illustration). 1 Rail The delivery package contains a cable clip which is used to guide the cable of the foot control panel away from the operating table.
  • Page 107 Description G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 108: S81 Ceiling Mount

    S81 ceiling mount Intended use The S81 ceiling mount is a suspension system for Zeiss surgical micro- scopes. It is used to power and control the motorized functions of the sur- gical microscope. The hallmarks of the S81 ceiling mount are its superb mobility and easy operation.
  • Page 109: Description Of The Modules

    Description Description of the modules The S81 ceiling mount comprises a column, a carrier arm and a suspen- sion arm. The suspension arm with the lamp housing and the control unit is mounted on the carrier arm. The control unit is rotatable through 180° (70° if the Superlux Eye illumination system with additional, integrated halogen illumination is used) and contains all electrical supply systems required for the control of a motorized surgical microscope.
  • Page 110: Design

    Description Design 1 Column 2 Carrier arm 3 Suspension arm 4 Illumination system, see page 72 5 Control panel G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 111 Description G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 112: Power Switch, Connector And Socket (Option)

    Description Power switch, connector and socket (option) The connector and socket can be either installed in the OR, or they can be integrated in the ceiling mount, at the back of the carrier arm (see illus- tration). 1 Rail The delivery package contains a cable clip which is used to guide the cable of the foot control panel away from the operating table.
  • Page 113 Description G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 114: Visu 160 Surgical Microscope On S88 Floor Stand

    Description VISU 160 surgical microscope on S88 floor stand Intended use The VISU 160 surgical microscope has been designed for the magnified visualization of the field of view during surgical procedures in ophthal- mology. The S88 floor stand powers and controls the motorized functions of the VISU 160 surgical microscope.
  • Page 115 Description G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 116: Visu 160 Surgical Microscope On S8 Ceiling Mount

    Description VISU 160 surgical microscope on S8 ceiling mount Intended use The VISU 160 surgical microscope has been designed for the magnified visualization of the field of view during surgical procedures in ophthal- mology. The S8 ceiling mount powers and controls the motorized functions of the VISU 160 surgical microscope.
  • Page 117 Description G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 118: Visu 160 Surgical Microscope On S81 Ceiling Mount

    Description VISU 160 surgical microscope on S81 ceiling mount Intended use The VISU 160 surgical microscope has been designed for the magnified visualization of the field of view during surgical procedures in ophthal- mology. The S81 ceiling mount powers and controls the motorized functions of the VISU 160 surgical microscope.
  • Page 119 Description G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 120: Foot Control Panel (Option)

    Description Foot control panel (option) Intended use The foot control panel permits you to control 14 different functions of a suspension system or surgical microscope, provided these functions are part of your configuration (suspension system, surgical microscope). The assignment of the functions to the controls of the foot control panel is shown on the next page.
  • Page 121 Description G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 122 Description The illustration shows the standard assignment of functions to the foot control panel. On request, our service staff can change the assignment of the focus/zoom functions and that of button (10). 1 Joystick for X-Y coupling 2 Reducing the lamp brightness 3 Recentering of the X-Y coupling and focus starting position (optional: control of Stereo Digital Inverter IIe / 3e from the company Oculus).
  • Page 123: Preparations

    Preparations Preparations Attaching the equipment Mounting the surgical microscope Mounting the tube, the eyepieces and the objective lens Changing the microscope accessories Connections Connecting the surgical microscope Connecting the S light guide Strain relief device on S88 floor stand Connecting the S88 floor stand Relocating the system Adjusting the supension system Adjusting the balance setting of the suspension arm...
  • Page 124: Attaching The Equipment

    Preparations Attaching the equipment Mounting the surgical microscope Warning! The maximum weight of the microscope including accessories must not exceed 20 kg! • Bring the suspension arm into its horizontal position, pull out locking knob (1) and turn it clockwise or counterclockwise through 180°. At the same time, slightly move the suspension arm up and down until the lock snaps in.
  • Page 125 Preparations G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 126 Preparations • Screw in securing screw (7) and tighten it firmly. Securing screw (7) must go into groove (9). This is ensured when the securing screw is flush with the outer surface. • Hook coupling (12) with the attached surgical microscope from above into receptacle (11) on the suspension arm, and tilt the coupling downward into its vertical position.
  • Page 127 Preparations G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 128: Mounting The Tube, The Eyepieces And The Objective Lens

    Preparations Mounting the tube, the eyepieces and the objective lens • Bring the suspension arm in a position convenient for you and firmly tighten locking knob (1). • Loosen securing screw (5) by a few turns. • Remove cover (3) and store it in a safe place. •...
  • Page 129 Preparations G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 130: Changing The Microscope Accessories

    Preparations Changing the microscope accessories You can change the microscope accessories in reverse order to that de- scribed before. Please observe the following: • Turn off the unit at the power switch before changing any accessories. • Bring the suspension arm in a position convenient for you and firmly tighten locking knob (1).
  • Page 131 Preparations G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 132: Connections

    Preparations Connections Connecting the surgical microscope • Turn locking cap (1) by a quarter turn to the right or left and pull up cov- er (2). • Plug microscope connector (3) into connector (4) and tighten the se- curing screws on the microscope connector. •...
  • Page 133 Preparations Surgical microscope, shown rotated G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 134: Strain Relief Device On S88 Floor Stand

    Preparations Strain relief device on S88 floor stand Note: To prevent inadvertent unplugging of the power cable and of the control element connector, secure the two cables in strain relief device (1). After strain relief device (1) has been mounted, the cables must have the following length: –...
  • Page 135 Preparations 320 mm G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 136: Connecting The S88 Floor Stand

    Preparations Connecting the S88 floor stand • Check the voltage indicated at (3). Caution: The voltage of the stand is factory-set to the rated voltage used in the country of destination. The rated voltage indicated at window (3) must cor- respond to the rated voltage available at the site of installation.
  • Page 137 Preparations G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 138: Relocating The System

    Caution: Over longer distances (e.g. removal, return for repair, etc), the instrument must always be transported in the original packaging or in special return packaging. For details, please contact your dealer or the Carl Zeiss service team. G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0...
  • Page 139 Preparations G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 140: Adjusting The Supension System

    Preparations Adjusting the supension system Adjusting the balance setting of the suspension arm • Only perform the balance setting procedure with the complete micro- scope equipment attached! Note: We recommend that you perform coarse balancing of the suspension arm before starting with precise balance setting of the suspension arm. The suspension arm must be locked in the horizontal position for this proce- dure.
  • Page 141 Preparations G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 142: Adjusting The Limit Of Downward Movement

    Preparations Adjusting the limit of downward movement The downward travel of the suspension arm must be limited in such a way that no hazard is caused to the patient's safety even if the surgical micro- scope is inadvertently lowered. • Give adjustment screw (1) a few turns to loosen it.
  • Page 143 Preparations G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 144: Positioning The S8 Ceiling Mount

    Preparations Positioning the S8 ceiling mount 1 Working position 2 Parking position Working position • Pull the ceiling mount into the working position using the handle (4). The recommended height is approx. 1750 mm measured from the handle to the floor. •...
  • Page 145 Preparations G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 146: Settings On The Control And Display Panel

    Preparations Settings on the control and display panel Adjusting the suspension system • Turn on the suspension system at its power switch. • Successively select the following functions on the suspension system: – lamp brightness, – motor speeds for zoom, –...
  • Page 147: Adjusting The Surgical Microscope

    Preparations Adjusting the surgical microscope Bring the surgical microscope into its starting position within the focusing range. Adjust the minimum magnification on the surgical microscope. Bring the surgical microscope into the position required. Adjust your interpupillary distance on the binocular tube. Adjust your prescription on the eyepieces.
  • Page 148: Adjusting The Tilt Angle

    Preparations Adjusting the tilt angle Using knob (1), you can position the surgical microscope in a range from +180° to -180° (+ in the direction of the surgeon and - in the opposite di- rection). The +90° setting is ideal for surgery on patients in a seated po- sition or lying on their side.
  • Page 149: Adjusting The Microscope Tilt To Angles Greater Than 15

    Preparations Adjusting the microscope tilt to angles greater than 15° • Remove all accessories mounted on the objective lens (e.g. 0° or 8° assistant's microscope). • Turn knob (1) until the surgical microscope is in the viewing position required. After the viewing angle has been set, the surgical microscope remains in this position.
  • Page 150 Preparations G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 151: Operation

    Operation Operation Checklist When using a wide-angle observation system (e.g. BIOM 3) Positioning the S88 floor stand Using the display and key field General functions Operating the OPMI® on the suspension system Procedure What to do in an emergency Failure of the halogen lamp Lamp failure in Superlux Eye illumination system Failure of lamp control Failure of focusing system...
  • Page 152: Checklist

    Operation Checklist Warning! If a function fails, you must not use this instrument for safety reasons. Cor- rect the fault (see the "Troubleshooting table“) or contact our service dept. Always check the following points before surgery (without patient!): – Check that the correct rated voltage has been set. –...
  • Page 153 Operation Suspension systems Note: After switching on, the suspension system automatically performs a self- test which takes approx. five seconds. The suspension system is equipped either with a halogen illumination system, a Superlux Eye illumination system or with a Superlux Eye illumi- nation system with additional, integrated halogen illumination (option).
  • Page 154 Operation – If operated for 8 hours/day on 5 days/week, the lamp needs to be ex- changed after 12 weeks maximum. Balance setting – Check that the suspension arm has been properly balanced. When the release button on the surgical microscope is pressed, the effort re- quired to move the arm up or down must be the same.
  • Page 155: When Using A Wide-Angle Observation System (E.g. Biom 3)

    Operation When using a wide-angle observation system (e.g. BIOM 3) When using a wide-angle observation system (e.g. BIOM 3 from Oculus) which is usually installed between the surgical microscope and the pa- tient, make sure that the patient is neither put at risk nor injured by the mo- torized focusing system or the movement of the suspension system arm.
  • Page 156: Positioning The S88 Floor Stand

    Operation Positioning the S88 floor stand Note: Please also read the chapter: "Relocating the stand", page 138. • Unlock brake tab (2). • Use maneuvering handle (1) to move the stand to the site of use. Make sure that movement is not obstructed by the power cord and the cable of the foot control panel.
  • Page 157 Operation G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 158: Using The Display And Key Field

    Operation Using the display and key field General functions Basic mode Operating keys "∇" and "∆" Halogen Keys "∇" and "∆" always refer to the display field (LCD) on the left, i.e. you can use them to change the values or settings currently displayed in this field.
  • Page 159 Operation "MANUAL" key The "MANUAL" key permits you to switch to manual operation. The mo- torized control functions of the surgical microscope are deactivated. The lamp brightness is automatically adjusted to a fixed setting, the value being shown in the first display. When the manual mode is activated, the yellow LED is lit and the word "MANUAL"...
  • Page 160 Operation Basic mode Basic mode The basic mode is always displayed in the normal operating status. Halogen In the basic mode, the following settings are displayed, depending on the installed surgical microscope: In the upper display field the current lamp brightness In the middle display Halogen: backup lamp field...
  • Page 161 Operation Lamp 1: xenon Lamp 2: halogen Acoustic signals Three succes- – Error message during the software check after sive beeps power-on of the suspension system. – Error message in the case of an internal system er- ror. One beep When the focus or zoom position is saved.
  • Page 162: Operating The Opmi On The Suspension System

    Operation ® Operating the OPMI on the suspension system Overview: user interface for the OPMI Basic mode Basic mode Speed mode FOCUS SPEED: USER Configuration ZOOM SPEED: mode 1 MODE STORE MANUAL SPEED: FOOT LEFT C: Configuration PHOTO mode 2 FOOT RIGHT D: MODE...
  • Page 163 Operation Control functions for the OPMI The control functions for the OPMI have been combined in 4 modes: Basic mode Speed mode – Setting the lamp brightness – Setting the speed for focusing – Setting the user ID – Setting the speed for the zoom function –...
  • Page 164 Operation OPMI: setting the lamp brightness This function permits you to set the lamp brightness. Basic mode In the basic mode, the lamp brightness currently set is shown in the upper display field. The middle display field remains empty and the associated keys "∇" and "∆"...
  • Page 165 Operation OPMI: Setting the adjustment speeds Speed mode You can set the adjustment speeds for the following microscope func- tions: – Focusing FOCUS SPEED: – Zoom function ZOOM – Adjustment of the X-Y coupling SPEED: Path: The basic mode is automatically displayed after the instrument has SPEED: been switched on.
  • Page 166 Operation Caution: Be extremely careful when changing these settings. You should change settings only under your own user ID. Notify all users of any changes, or make sure that each user only works under his own user ID. Path: To return to the basic mode, press the "MODE" key in the speed mode or do not press any key for 20 seconds.
  • Page 167 Operation OPMI: Assigning a function to buttons C and D of the foot control panel Buttons C and D of the foot control panel can be configured as required. In configuration mode 1, the upper and middle display and key fields permit you to assign one of the following functions to buttons C and D: Display Function of button C or D of the foot control panel...
  • Page 168 Operation Roll-over procedure Use the "Roll-over procedure" to select the required function in the default sequence of the above table. Every time you press the relevant button "∇" you advance clockwise. Every time you press the relevant button "∆" you advance counterclock- wise.
  • Page 169 Operation FOC-MEM To save a new focus position, press the key configured for this purpose for longer than 2 seconds. A beep is emitted when the new position has been saved. To move to a stored focus position, press the key config- ured for this purpose only briefly (less than 2 seconds).
  • Page 170 Operation OPMI: Setting the adjustment speeds of focus and X-Y coupling as a function of the zoom setting Configuration In this mode, you can select dynamic speed control for focusing and the mode 1 X-Y coupling. The depth of field of the optical system changes as a function of the zoom setting: FOOT LEFT C:...
  • Page 171 Operation Path: After you have switched on the instrument, the basic mode is auto- matically displayed. To access configuration mode 1 from the basic mode, simultane- ously press the "MODE" and "STORE" keys. Rollover Select the function required using a rollover run in the sequence defined in the above table.
  • Page 172 Operation ® OPMI : Selecting the magnetic brakes to be unlocked You can define which magnetic brakes of the suspension system are to be unlocked by turning the left and/or right handgrips (1) and (2) (turning function). In configuration mode 2, the upper display and key field (SELECT BRAKES) permits you to select the following: XY - Z The magnetic brakes of the suspension system for X-Y move- ment and Z movement can be unlocked separately.
  • Page 173 Operation Path: To return to the basic mode, press the "MODE" key in the configu- ration mode or do not press any key for 20 seconds. To access the next configuration mode 3, simultaneously press the "MODE" and "STORE" keys. G-30-1529-en OPMI®...
  • Page 174: Procedure

    Operation Procedure Caution! – Avoid looking directly into the light source, e.g. into the microscope ob- jective lens or into the light guide! – When selecting the brightness level for the patient's eye, always take care to keep the strain on the patient's eye to a minimum. –...
  • Page 175 Operation • Select the magnification required (zoom). Look through the eyepieces of the binocular tube. Adjust the eyepieces in such a way that you can see both the edge of the field of view and the microscope image sharply. Also see “Adjusting the surgical microscope“. •...
  • Page 176: What To Do In An Emergency

    Operation What to do in an emergency Failure of the halogen lamp Caution: Do not cover the ventilation grid (2)! Make sure that drapes do not cover the grid. This can lead to overheating of the lamp modules and to lamp failure.
  • Page 177 Operation G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 178: Lamp Failure In Superlux Eye Illumination System

    Operation Lamp failure in Superlux Eye illumination system Caution: Do not cover the ventilation grid! For example, drapes could be covering the grid. This can lead to overheating of the lamp module and to lamp failure. Note: Yellow indicator lamp (4) lights when the lamp has failed, or if the lamp module is defective.
  • Page 179 Operation G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 180: Failure Of Lamp Control

    Operation Failure of lamp control • Press the Manual key (1) if brightness control is no longer possible. Note: When the manual function has been activated, all electrical control sys- tems are inoperative. The lamp brightness is automatically adjusted to a fixed setting.
  • Page 181 Operation G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 182: Failure Of Magnetic Brakes

    Operation Failure of magnetic brakes If the magnetic brakes fail (magnetic brakes are locked), you can manu- ally position the articulated arm including the microscope by overcoming the locking effect of the magnetic brakes. Failure of the X-Y coupling • Disconnect the surgical microscope from the suspension system.
  • Page 183: Failure Of The Zoom Function

    Operation Failure of the zoom function • Disconnect the surgical microscope from the suspension system. The connector is located under the cover of the suspension arm. The illu- mination remains on. • Use zoom adjustment knob (1) of the microscope to manually set the magnification required (if necessary, use a tool, e.g.
  • Page 184 Operation G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 185: Maintenance / Further Information

    Maintenance / Further information Maintenance / Further information Trouble-shooting Changing the halogen lamp Changing the Superlux Eye xenon lamp module Magnifications / Fields of view Care of the unit Sterilization Disinfecting the control keys Ordering data Spare parts Accessories Disposal G-30-1529-en OPMI®...
  • Page 186: Trouble-Shooting

    Maintenance / Further information Trouble-shooting • This instrument is a high-grade technological product. To ensure opti- mum performance and safe working order of the instrument, its safety must be checked once every 12 months. We recommend having this check performed by our service representative as part of regular main- tenance work.
  • Page 187 Maintenance / Further information Problem Possible cause Remedy Surgical field illumination Brightness level set too Adjust brightness using the too bright. high. control on the suspension system or the foot control panel. Switch off illumination on suspension system. Illuminate surgical field using an OR illuminator.
  • Page 188 Maintenance / Further information Problem Possible cause Remedy Surgical field illumination on Thermal cut-out activated. Remove the cause of over- microscope not working. heating. example, drapes could be covering the grid. When the lamp module has cooled down, the illumination switches on again.
  • Page 189 Maintenance / Further information Problem Possible cause Remedy Motorized focusing Manual function is acti- Switch off manual function. zoom functions of surgical vated. (Yellow LED above microscope are inoperative. the key lights up). Suspension arm is in hori- Suspension arm still locked. Pull out locking device and zontal position and cannot turn though 180°.
  • Page 190 Maintenance / Further information Halogen illumination system Problem Possible cause Remedy Surgical field illumination on Thermal cut-out activated. Remove the cause of over- microscope not working. heating. example, drapes could be covering the grid. When the lamp module has cooled down, the illumination switches on again.
  • Page 191 Maintenance / Further information Problem Possible cause Remedy Surgical field illumination Brightness level set too Adjust brightness using the page 146 too bright. high. control on the suspension system or the foot control panel. Switch off illumination on suspension system. Illuminate surgical field...
  • Page 192 Maintenance / Further information Superlux Eye illumination system Problem Possible cause Remedy Surgical field illumination on Thermal cut-out activated. Remove the cause of over- microscope not working. heating. example, drapes could be covering the grid. When the lamp module has cooled down, the illumination switches on again.
  • Page 193 Maintenance / Further information Problem Possible cause Remedy No surgical field illumina- Xenon lamp does not ignite. Switch off illumination on tion. suspension system. Illuminate surgical field using an OR illuminator. Contact service dept. Xenon lamp is lit, but beep Defective lamp control...
  • Page 194: Changing The Halogen Lamp

    Maintenance / Further information Changing the halogen lamp Warning! If you change the lamp shortly after it has failed, the lamp will still be very hot. Wear heat-protection gloves to avoid burns! Note: If both halogen lamps (4 and 5) fail during surgery, insert an available backup lamp module.
  • Page 195 Maintenance / Further information G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 196 Maintenance / Further information • Plug the ceramic base (8) onto the contact pins of the new halogen lamp. • Insert the new halogen lamp. Make sure you do not touch the lamp bulb (7) or the interior of the reflector (6). •...
  • Page 197 Maintenance / Further information G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 198: Changing The Superlux Eye Xenon Lamp Module

    Check the function of the xenon lamp and backup xenon lamp. Pack the old lamp module (5) in the transport package of the new lamp module. Fill in the enclosed return card and send the old lamp module to the nearest Carl Zeiss service agency. Note: •...
  • Page 199 Maintenance / Further information G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 200: Magnifications / Fields Of View

    Maintenance / Further information Magnifications / Fields of view Using the magnification factor γ of the zoom system, you can calculate the total magnification of the surgical microscope according to the following formula: tube ---------- - γ M ⋅ ⋅ where: is the focal length of the binocular tube tube...
  • Page 201: Care Of The Unit

    Remove minor contaminations such as dust, streaks, etc. using a clean microfiber cleaning cloth (available from specialized dealers or under Cat.No. 1254-655). Auxiliaries from Zeiss: optics cleaning set, Cat. No. 1216-071 Suitable for the cleaning of objective lenses and eyepieces of the surgical microscope at regular intervals.
  • Page 202: Sterilization

    Wipe off any residue with a mixture of 50% ethyl alcohol and 50% distilled water plus a dash of household dish-washing liquid. Sterilization The asepsis sets available from Carl Zeiss contain rubber caps, sleeves and handgrips which can be sterilized in autoclaves. We recommend the following program for sterilization: Sterilization temperature: 134°...
  • Page 203: Disinfecting The Control Keys

    To be able to use the system in the OR, for example, it may be necessary to disinfect the control keys. We recommend using MELISEPTOL disin- fectant solution (B. Braun, Melsungen AG). Carl Zeiss keeps MELI- SEPTOL in stock, and you can also obtain it locally in many countries from representatives of B.
  • Page 204: Ordering Data

    Maintenance / Further information Ordering data VISU 160 surgical microscope Description Cat. No. VISU 160 surgical microscope including X-Y coupling 30 26 05- 9001 12.5x eyepiece (2x) 30 55 43- 9901 45° inclined tube (option) 30 37 84- 0000 Invertertube™ - 110° tiltable tube 30 37 97- 9120 180°...
  • Page 205 Maintenance / Further information S8 ceiling mount Description Cat. No. S8 ceiling mount 1176-968 Coupling for VISU 30 59 52- 8030 Halogen illumination system, one module 1174-210 Halogen illumination system, two modules 1174-211 Superlux Eye illumination system 30 49 77- 9010 S light guide, 2.0 m 30 34 81- 9020 Complete replacement lamp module with 2 xenon...
  • Page 206: Spare Parts

    Maintenance / Further information Upgrade kit for Superlux Eye, for retrofitting an 30 49 77- 9022 existing xenon illumination system with addi- tional, integrated halogen illumination (option) Spare parts VISU 160 surgical microscope Description Cat. No. Asepsis set, 12 mm (pack of 6) 305810-9002-000 Asepsis set, 22 mm (pack of 6) 305810-9001-000...
  • Page 207 Maintenance / Further information Suspension systems - Superlux Eye illumination system Superlux Eye illumination system Description Cat. No. Complete replacement lamp module with 2 xenon 304977-9036-700 lamps in transport container and with return card; in exchange for a returned module with defective xenon lamps Complete xenon lamp module with 2 xenon lamps 304977-9036-000...
  • Page 208: Accessories

    Please observe the following: Only operate the instrument with the accessories included in the delivery package. If you want to use other accessories, make sure that Carl Zeiss or the manufacturer of the accessories has proved and confirmed that these accessories meet the respective technical safety standards and can be used without risk.
  • Page 209: Disposal

    Maintenance / Further information Disposal User information on the disposal of electrical and electronic devices This symbol means that the product must not be disposed of as normal domestic waste. The correct disposal of electrical or electronic devices helps to protect the environment and to prevent potential hazards to the environment and/or human health which may occur as a result of improper handling of the de- vices concerned.
  • Page 210 Maintenance / Further information G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 211: Technical Data

    Technical data Technical data Technical data Ambient requirements CE conformity Changes to the system G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 212 Technical data Technical data VISU 160 surgical microscope Optical data Magnification 4.3x - 25.5x (with f = 200 mm and 12.5x widefield eyepieces) Motorized zoom system with apochromatic op- tics, 1:6 zoom ratio, magnification factor γ = 0.4x - 2.4x. Focusing Motorized, focusing range 50 mm At the press of a button, the focusing drive moves...
  • Page 213 Technical data Swing-in stops Patented retinal protection device Pat. No. DE 33 39 17 2 C2 Patented field stop for reduced glare from the sclera Pat. No. G 91 03 43 3.7 Horizontal slit, 2.5 mm wide, continuously mov- able in the vertical direction in the field of view Horizontal slit, 5 mm wide, continuously movable in the vertical direction in the field of view Vertical slit, 2.5 mm wide, snaps in at the center...
  • Page 214 Technical data S88 floor stand Mechanical data Suspension arm Length...850 mm Swivel angle...320° Vertical lift...± 360 mm Carrier arm Length...450 mm Swivel angle...320° Stand height 1880 mm Base 805 x 805 mm (length x width) Admissible max. 20 kg (complete microscope equipment, in- load on suspension cluding accessories) Total weight...
  • Page 215 Technical data 270° G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 216 Technical data Electrical data of S88 floor stand Power requirements Only connect the suspension system to wall out- lets which are provided with a properly connected protective ground conductor. Rated voltage 115 VAC (100...125 VAC± 10%) 230 VAC (220...240 VAC± 10%) Current consump- 115 VAC max.10 A...
  • Page 217 Technical data Halogen illumination system Fiber optic illumina- Lamp housing with 2 halogen reflector lamps tion (1 backup lamp) with 12 V 100 W in quick-change modules for one light guide, with GG 475 (retina protection filter) and KK40 filter (to increase the color temperature), fully automatic lamp change when the first hal- ogen lamp fails.
  • Page 218 Technical data S8 ceiling mount Mechanical data Suspension arm Length...930 mm Swivel angle...270° Vertical lift...± 360 mm Swivel angle of control panel...180° (± 90°) * Xenon illumination system with additional inte- grated halogen illumination (option) Swivel angle of control panel...± 35° Lift and carrier arms Length...1000 mm Swivel angle...270°...
  • Page 219 Technical data 1000 mm max. 930 mm max. 700 mm 270° ±90° max. 350° (* ±35°) 360 mm ca. 1750 mm 270° 360 mm G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 220 Technical data Electrical data of S8 ceiling mount Rated voltage 230 VAC (220...240 VAC± 10%) 115 VAC (100...125 VAC± 10%) Current 230 VAC 3 A consumption 120 VAC 5 A 100 VAC 6 A Rated frequency 50...60 Hz Fuses Automatic circuit breaker Remote socket Electrical –...
  • Page 221 Technical data Halogen illumination system Fiber optic illumina- Lamp housing with 2 halogen reflector lamps tion (1 backup lamp) with 12 V 100 W in quick-change modules for one light guide, with GG 475 (retina protection filter) and KK40 filter (to increase the color temperature), fully automatic lamp change when the first hal- ogen lamp fails.
  • Page 222 Technical data S81 ceiling mount Mechanical data Suspension arm Length...930 mm Swivel angle...270° Vertical lift...± 360 mm Swivel angle of control panel...180° (± 90°) * Xenon illumination system with additional inte- grated halogen illumination (option) Swivel angle of control panel...± 35° Carrier arm Length...600 mm Swivel angle...270°...
  • Page 223 Technical data 600 mm max. 930 mm max. 350° 270° 360 mm 270° ±90° (* ±35°) 360 mm G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 224 Technical data Electrical data of S81 ceiling mount Rated voltage 230 VAC (220...240 VAC± 10%) 115 VAC (100...125 VAC± 10%) Current consump- 230 VAC 3 A tion 120 VAC 5 A 100 VAC 6 A Rated frequency 50...60 Hz Fuses Automatic circuit breaker Remote socket View of connector side...
  • Page 225 Technical data Halogen illumination system Fiber optic illumina- Lamp housing with 2 halogen reflector lamps tion (1 backup lamp) with 12 V 100 W in quick-change modules for one light guide, with GG 475 (retina protection filter) and KK40 filter (to increase the color temperature), fully automatic lamp change when the first hal- ogen lamp fails.
  • Page 226 Technical data Ambient requirements For operation Temperature +10 °C...+40 °C Rel. humidity 30%...75% Air pressure 700 hPa...1,060 hPa For transportation and Temperature - 40 °C...+70 °C storage Rel. humidity 10%...100% (without condensation) Air pressure 500 hPa...1,060 hPa CE conformity The system meets the essential requirements stipulated in Annex I to the 93/42/EEC Directive governing Medial Devic- es.
  • Page 227: Index

    Anti-fogging agent Aperture selector 216, 220, 224 Approval Arrows indicating the focusing range AUX signal 98, 167 Auxiliaries from Zeiss Backup lamp, switching to Balance setting screw Basic mode 76, 80, 86 Brightness control Brightness of illumination Cable and light guide clip...
  • Page 228 Index CE conformity CE label Ceiling anchors Changes to the system Changing the halogen lamp. Checklist Cleaning method, recommended Cleaning optical surfaces Cleaning, mechanical surfaces Collision and crushing, risk of Composition, spectral Connecting the S light guide Connecting the surgical microscope Connection panel Connections 106, 112...
  • Page 229 Index 74, 84 Filter selector knobs FOC-MEM Focus memory Focus of the light source Focus speed Focusing system failure Fogging of optical surfaces Foot control panel Foot control panel, buttons C and D 36, 38, 74, 84 GG 475 filter GG 475 protection filter 76, 80, 86 Green indicator lamp...
  • Page 230 Index 176, 178 Lamp failure 74, 78, 83 Lamp module LB value LCD display LCD display field Light guide socket Light source, focus Lock of the cable duct Locking and adjustment screws Magnetic brake failure Magnetic brakes, selection of Magnifications Maintenance 82, 84 Manual activation of backup lamp...
  • Page 231 Index PHOTO Photo Phototoxic retinal injury in eye surgery Potential equalization Power outlet Power switch 98, 106, 112 Preparations Procedure Protective ground contact Receptacles Re-equipping the instrument Release bar Relocating the system 98, 112 Remote control socket Repeat function Requirements for operation Retina protection filter GG 475 Risk of fire Row of keys, operation...
  • Page 232 Index Support arm with tilt device Surgical microscope Surgical microscope on suspension system 211, 212 Technical data Time of exposure to light Trouble-shooting table Tube and eyepieces for main microscope Tube, eyepieces and objective lens, mounting Use of the instrument, before Use of the instrument, for every Use, immediately after Use, notes...
  • Page 233 G-30-1529-en OPMI® VISU 160 on S8, S81 & S88 Suspension Systems Issue 5.0 Printed on 09. 10. 2006...
  • Page 234 Carl Zeiss Surgical GmbH Fax: +49 (0) 7364 20-4823 E-mail: surgical@zeiss.de 73446 Oberkochen www.zeiss.com/surgical Germany G-30-1529-en Printed in Germany CG-ST...