90
Adjusting the limit of downward movement
G-30-1426-en
Operation
The vertical lift of the suspension arm must be limited in such a way that
the patient's safety is also ensured when the microscope is inadvertently
lowered.
•
Loosen the adjustment screw (1) by a few turns.
•
Press one of the release keys for the magnetic brakes on the surgical
microscope and lower the microscope to a position where it can be fo-
cused on the surgical field (depending on the focal length of the objec-
tive lens) and where the safety distance to the surgical field is still
sufficient.
•
Turn the adjustment screw (1) clockwise as far as it will go.
•
Move the surgical microscope again to the lower limit and check the
safety distance.
S88 Floor Stand with Lifting Column
Issue 9.0
Printed on 28. 01. 2009