24
G-30-1426-en
Safety
Xenon illumination system
Warning!
The xenon lamp has a limited service life of 500 h.
If used beyond its maximum service life, the xenon lamp may explode.
•
Replace the xenon lamp in good time.
•
Reset the service hour counter to "0" after replacing the lamp.
Warning!
Lamp rupture (audible as a loud bang) may lead to jamming of the lamp
module and/or failure of the electronics modules.
•
Before opening the lamp housing, make sure that the system is moved
to a position where neither the patient nor the user is put at risk by fall-
ing items.
•
Do not continue using the system if the lamp module is jammed or the
illumination is no longer operational due to defective electronics mod-
ules. Inform our service department.
1 Switching to the backup lamp
The lamp module contains two xenon lamps. The second bulb is used
as a backup bulb which must be swung into the illumination beam path
if the first bulb fails.
•
If the xenon bulb fails, open the lamp module as follows: Press
button (7). The lamp module is slightly ejected.
•
Pull out the lamp module as far as it will go.
•
Turn knob (1) through 180° until it snaps in place. This moves the
backup lamp into the illumination beam path.
•
Push the lamp module all the way back into the lamp housing.
•
Reset service hour counter (6) to "0". Use a pointed object and press
it into the recess of reset button (5).
S88 Floor Stand with Lifting Column
Issue 9.0
Printed on 28. 01. 2009