Safety
Xenon illumination system
Warning!
The xenon lamp has a limited service life of 500 h.
If used beyond its maximum service life, the xenon lamp may explode.
Replace the xenon lamp in good time.
Reset the service hour counter to "0" after replacing the lamp.
Warning!
Lamp rupture (audible as a loud bang) may lead to jamming of the lamp
module and/or failure of the electronics modules.
Before opening the lamp housing, make sure that the system is moved
to a position where neither the patient nor the user is put at risk by fall-
ing items.
Do not continue using the system if the lamp module is jammed or the
illumination is no longer operational due to defective electronics mod-
ules. Inform our service department.
1 Switching to the backup lamp
The lamp module contains two xenon lamps. The second bulb is used
as a backup bulb which must be swung into the illumination beam path
if the first bulb fails.
If the xenon bulb fails, open the lamp module as follows: Press
button (7). The lamp module is slightly ejected.
Pull out the lamp module as far as it will go.
Turn knob (1) through 180° until it snaps in place. This moves the
backup lamp into the illumination beam path.
Push the lamp module all the way back into the lamp housing.
Reset service hour counter (6) to "0". Use a pointed object and press
it into the recess of reset button (5).