Changing The Xenon Lamp Module - Zeiss S88 Instructions For Use Manual

Floor stand with lifting column
Table of Contents

Advertisement

150

Changing the xenon lamp module

G-30-1426-en
Maintenance / Further information
Warning!
The lamp module must only be changed by appropriately trained per-
sonnel. Improper handling of the xenon lamp may lead to damage or in-
jury.
Warning!
Lamp rupture (audible as a loud bang) may lead to jamming of the lamp
module and/or failure of the electronics modules.
Before opening the lamp housing, make sure that the system is moved
to a position where neither the patient nor the user is put at risk by fall-
ing items.
Do not continue using the system if the lamp module is jammed or the
illumination is no longer operational due to defective electronics mod-
ules. Inform our service department.
Change the lamp module as follows:
First switch off the suspension system at the power switch.
Only change the lamp module after it has cooled down completely! In
the event of a malfunction, there is a risk of explosion due to the high
pressure inside the hot lamp. The hot surface of the xenon lamp may
also cause burns.
Only change the module after switching off the system. The igniter
produces high voltage when the system is on.
Press button (1). The lamp module is slightly ejected.
Pull out the lamp module as far as it will go.
Slide the original transport case (2) over the module, making sure that
bolt (3) engages in bore (4). This unlocks the stop.
Remove the old module and install the new lamp module by proceed-
ing in the reverse order.
Check the function of the xenon lamp and backup xenon lamp.
S88 Floor Stand with Lifting Column
Issue 9.0
Printed on 28. 01. 2009