Zeiss S88 Instructions For Use Manual page 22

Floor stand with lifting column
Table of Contents

Advertisement

22
G-30-1426-en
Safety
Stand lifting column
1 Key
to set the optimum viewing height of the surgical microscope or for
downward movement into the transport position.
As long as you keep switch (1) pressed down, the lifting column (2) in
the stand base moves upwards or downwards, depending on the po-
sition of the switch. When you release the switch, the lifting column
stops immediately.
– Before raising or lowering the suspension system, make sure that
there is sufficient clearance from other objects so that any collision is
avoided.
Warning!
– Do not activate the lifting column during surgery!
– Do not use the lifting column for focusing.
– When using the fundus imaging system which is installed between the
surgical microscope and the patient, make sure that the patient is nei-
ther put at risk nor injured
– by the motorized movement of the lifting column,
– by the motorized focusing system,
– by the movement of the suspension arm.
S88 Floor Stand with Lifting Column
Issue 9.0
Printed on 28. 01. 2009