Preparations for use
Warning!
It may happen that no second lamp module such as the VISULUX™ fiber
slit lamp is used on the microscope.
To prevent light guide (3) provided for this purpose from injuring the pa-
tient, you must attach it to the cable holder of the suspension system.
Switch off the illumination system for the unneeded light guide by set-
ting the illumination selector switch (4) to the left.
Press the light guide into an empty position of cable holder (2).
Remove cap (1) from the light guide. This prevents melting of the cap
in case the second illuminator is switched on by mistake.