34

Failure of all control functions (Emergency mode)

1
2
3
G-30-1702-en
Check whether power switch (3) has been switched on.
Remove magnetic plate (1) which covers emergency mode switch (2).
Set switch (2) from position 1 to position 2 (emergency mode).
The microscope and suspension system functions have now been com-
pletely deactivated. The emergency mode ensures that the illumination
continues to function with constant and sufficient intensity.
OPMI® Pentero® Software Release 2.2
Issue 2.0
Printed on 17. 03. 2008