8-8 Setting the Output Conditions

Setting the Filter

Setting Filters p.115
Setting the Median Filter
Setting the Median Filter p.116
As an example, here is an explanation of the procedure for selecting the "5" median filter setting.
Operating procedure
Steps
Key operation
1
2
3
4
5
6
7
ZW-7000 User's Manual
Display
Description
Press and hold the
mode.
Press the
Press the
key.
Press the
key.
Select "MEDIAN" as the filter type.
Press the
key.
Select the filter mode from the following:
OFF / 3 / 5 / 9 / 15 / 31
The current setting value is displayed on the sub-display.
Press
In this example, select "5" and press the
Press the
key for two seconds to enter the FUNC
/
key to select MEAS, then press the
/
keys to select "TASK1" and press the
/
keys to select "FILTER" and press the
/
keys to select "MEDIAN" and press the
/
key to enter editing mode and the sub-display blinks.
key for two seconds to enter the RUN mode.

Setting the Output Conditions

key.
key.
203