4-3 Setting the Output Conditions

Setting Scaling

This setting is used when you want to correct any errors that are generated due to the installation status of the
Sensor Head, and display the corrected value on the main display as a measured value.
There are three types of setting: "auto scaling" ("1-point scaling" and "2-point scaling") that automatically sets
the correction value of a placed sensing object, "fixed scaling" that manually sets the correction value and
"thickness scaling" that automatically sets the correction value of a placed sensing object (transparent object).
Important
The "Zero reset" p.92 settings return to the default settings when scaling is set. If scaling is set, perform a zero reset
as necessary.
Performing Manual Scaling
Set the scaling by entering the correction value.
Item
Setting item
Scaling
Scaling
(Manual scaling)
Span
Offset
130

Setting the Output Conditions

Setting value
OFF (default value)/ON
-10.0000 to 10.0000
(default value: 1.0000)
-999.999999 to 999.999999
(default value: 0)
Description
Set scaling ON/OFF.
Set the inclination of the sensor characteristics as a
coefficient.
Measurement value
(mm)
10.0
-10.0
A fixed value is added to/subtracted from the measured
value.
Measurement value
(mm)
Workpiece
displacement
Workpiece
displacement
ZW-7000/5000
User's Manual