Performing 2-point Scaling
Measurement is performed at two positions, and offset values are set for those measured values.
Example: When correcting display values to match actual distances
Display value
(mm)
4.0
3.8
-3.0
Point 1
Item
Setting item
Actual scaling
Set value
(two points)
(1st point)
Set value
(2nd point)
Important
Separate the two specified points by at least 1% of the rated measuring range for the connected Sensor Head.
For example, for the ZW-S7010, the two measured points must be separated by at least 1 mm × 0.01 = 0.01 mm as
the measuring range is 1 mm (± 0.5 mm).
 Explorer pane
→ Edit pane
→ Task Settings window : [Scaling]
1
Set the [Scaling] to OFF.
2
Click [Auto setting].
The [Scaling] popup menu appears.
112
Setting the Output Conditions
Point 2
Measured value
(mm)
4.0
-2.8
-3.0
Setting value
-999.999999 to 999.999999 [mm]
-999.999999 to 999.999999 [mm]
: [Bank] | [(Bank Data Name)] (double click)
: [Task Settings] icon (
Description
Correct the measured value of the 1st point, and set the
value to display.
Correct the measured value of the 2nd point, and set the
value to display.
)
ZW-7000 User's Manual