8-10 Setting Threshold Value
3-6 Setting Threshold Value p.92
There are two ways of setting the threshold value.
Method
Operating key
TEACHING
Direct
/
Teaching
Perform the measurement and set the result as the threshold value.
Sensor head
Threshold sample
(upper limit)
The existing settings for hold, trigger, and scaling are also reflected in the measurement during teaching.
As an example, here is an explanation of the procedure for setting the current measured value as a HIGH
threshold value.
ZW-7000 User's Manual
Description
Perform the measurement and set the result as the threshold value.
This is a convenient way when the upper or lower limit sample is available.
Set the threshold value by entering the numerical values directly.
/
/
This is a convenient way when you know the dimensions of a quality product or
when adjusting the threshold value set by teaching.
Teaching point
= HIGH threshold value
Sensor head
Teaching point
= LOW threshold value
Threshold sample
(lower limit)
Setting Threshold Value
225