Omron ZW-XC7 User Manual page 120

Confocal fiber type displacement sensor
Table of Contents

Advertisement

Setting the Frequency Filter
Set a filter to ignore or detect the changes in a specific frequency in the measured data.
Item
Setting item
Frequency filter
Frequency filter
Cutoff frequency
Cutoff frequency
(Upper limit)
Cutoff frequency
(Lower limit)
 Explorer pane
→ Edit pane
→ Task Settings window : [Filter]
1
Select the frequency filter type from
[Frequency filter].
2
Set the cutoff frequency from [Cutoff
frequency]. When a band pass filter is
selected, set [Cutoff frequency (upper limit)]
and [Cutoff frequency (lower limit)].
118
Setting the Output Conditions
Setting value
OFF (default value)
Lowpass filter:
0.040 to 23735.000
(default value: 0.040)
Highpass filter:
3.000 to 23725.000
(default value: 23725.000)
Bandpass filter
Lowpass filter:
1 to 23725[Hz] (default
value:2[Hz])
Highpass filter:
64 to 23725[Hz] (default
value:23725[Hz])
64 to 23725[Hz]
(default value:
Upper limit: 23725.000[Hz]
Lower limit:
128[Hz])
: [Bank] | [(Bank Data Name)] (double click)
: [Task Settings] icon (
Description
Frequency filter is not used.
Ignores frequency components larger than the
specified cut-off frequency. (Only gradual changes
are captured.)
Detects frequency components smaller than the
specified cut-off frequency. (Sudden changes are
captured.)
Detects the frequency components between the cut-
off frequency (lower limit) and the cut-off frequency
(upper limit).
Set this item when a low pass filter or a high pass
filter is selected.
Set this item when a band pass filter is selected.
)